User's Manual

1760 Power Quality Recorder
Preparing to Use the Recorder
7
emv03.bmp
Figure 3. Battery Label
XWWarning
Connect the supply cable of the device only
to sections CAT I, II or III of the supply system
(For details refer to the Identifications section)
and the voltage to earth should not exceed
300 V.
Input Voltage – Measuring Inputs
The measurement category (refer to the “Identifications”
section) and the maximum voltage to earth ground from
the sensors must conform to the power supply system.
Refer to “Inscription” and “Technical Specifications”
section.
Protection
IP40
Preparing to Use the Recorder
Before you use the Recorder, carefully inspect the
packaging and contents to ensure that there is no damage.
Note
Depending on the configuration you ordered,
flexible current probes and voltage sensors may
not be part of the standard equipment. An
internal transient capture option (factory installed
only) is another configuration that is determined
at time of order.