User's Manual

PQ Log
Users Manual
28
Number of events
The available messages are:
No job is programmed
Job is active (between segments)
Job is active, previous job is available
Job is completed
Additionally the following information is also available:
Logger name
Logger date/time
Hardware type of the Logger
Serial number
The field ”Current sensor type” shows the type of the connected accessory:
o Type 0: No accessory detected
o Type 1, 3: Flexi Set 3-phase
o Type 2, 4: Flexi Set 3-phase+N
o Type 5: 3-phase current clamp set
o Type 6: Current clamp set 3-phase + N
o Type 7: 1-phase clamp set
o Type 8: 1-phase Flexi Set
Note
Types 1 and 2 are sensors which were shipped with the earlier instruments
that are compatible with the Logger.
Max. Clamp current
Shows the nominal measuring range of the actual accessory.
Logger Details
This screen displays type of analog and digital board, version of Boot-, CPU and DSP
software.
Online Data from Logger
Starts the ONLINE function
as previously described. This is the stripchart of live
readings. This feature can be used only when a logger has been set up and logging is
active.
.
Analysis
This menu option enables the analysis of a Logger data file:
Note
This menu is available only if a measurement file is open.