Manual Supplement Manual Title: 1745 Users Part Number: 2560366 Print Date: April 2006 Revision/Date: 1, 7/06 Supplement Issue: 6 Issue Date: 4/14 Page Count: 5 This supplement contains information necessary to ensure the accuracy of the above manual. © 2008-2014 Fluke Corporation. All rights reserved.
1745 Users Manual Supplement Change #1 On page 4, add the following symbol to Table 1: Do not apply around or remove from the Hazardous Live conductors. On page 7, replace Table 3 with: Table 3.
Manual Supplement 1745 Users Change #2 On page 16, under Connecting the Logger add the following text and Figure: To avoid shock or personal injury, keep fingers behind the tactile barrier, see the figure. Single Insulated Current Carrying Conductor AC 60 Release Button C U 0 V R R EN C AT SE R IA L T III i5 C s LA N M U P M BE R Load Direction Arrow Tactile Barrier cab01f.
1745 Users Manual Supplement Change #4 On page 7, Table 2, change the following: From: Shielded 4 Phase Flexi Set (15 A/150 A/1500 A/3000 A) FS17XX FS17X5-TF-II, 3000A 4-PH FLUKE 1735 & 1745 THIN FLEXI SET 4531791 To: Add the following to Table 2: VL1735/45, Banana 4-Phase Voltage Lead Set for Fluke-1735/45 3276205 Change #5, 57885 On page 45, under Environmental Specifications, following Safety add: Altitude 4/14 2000 m 3
1745 Users Manual Supplement Change #6 On page 18, following the Note, add: Operating Instructions 1. Connect the Probe to the 1745 via the current input socket. 2. Wear protective equipment or de-energize the circuit and place the Probe around the conductor under test. To lock the coil, see the Figure. 3. Re-energize the circuit. 4. Observe and take measurements as required. Positive output indicates that the current flow is in the direction shown by the arrow on the Probe. 5.
1745 Users Manual Supplement IP Rating IEC 60529: IP50 Safety: IEC 61010-1: Overvoltage 600 V Category III, 300 V Category IV, Pollution Degree 2 IEC 61010-2-030: Measurement 600 V CAT III, 300 V CAT IV IEC 61010-2-031: Measurement 600 V CAT III, 300 V CAT IV IEC 61010-2-032: Measurement 600 V CAT III, 300 V CAT IV Electromagnetic Compatibility Radio Frequency Emissions Applies to use in Korea only.