550C/1555 Insulation Tester Users Manual PN 3593019 April 2010 © 2010 Fluke Corporation. All rights reserved. Printed in USA. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
LIMITED WARRANTY AND LIMITATION OF LIABILITY Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is three years and begins on the date of shipment. Parts, product repairs, and services are warranted for 90 days.
LIMITES DE GARANTIE ET DE RESPONSABILITE La société Fluke garantit l’absence de vices de matériaux et de fabrication de ses produits dans des conditions normales d’utilisation et d’entretien. La période de garantie est de trois ans et prend effet à la date d’expédition. Les pièces, les réparations de produit et les services sont garantis pour un période de 90 jours.
BEGRENZTE GEWÄHRLEISTUNG UND HAFTUNGSBESCHRÄNKUNG Fluke gewährleistet, daß jedes Fluke-Produkt unter normalem Gebrauch und Service frei von Materialund Fertigungsdefekten ist. Die Garantiedauer beträgt 3 Jahre ab Versanddatum. Die Garantiedauer für Teile, Produktreparaturen und Service beträgt 90 Tage. Diese Garantie wird ausschließlich dem Ersterwerber bzw.
GARANZIA LIMITATA & LIMITAZIONE DI RESPONSABILITÀ Ogni prodotto Fluke è garantito come esente da difetti nei materiali e nella manodopera per normali situazioni di uso. II periodo di garanzia è di tre anni a partire dalla data di spedizione. La garanzia per le parti sostituite, le riparazioni e l’assistenza è di 90 giorni. La garanzia è emessa solo a beneficio dell’acquirente originale o del consumatore finale che abbia acquistato il prodotto da un rivenditore Fluke autorizzato.
GARANTÍA LIMITADA Y LIMITACIÓN DE RESPONSABILIDAD Se garantiza que todo producto de Fluke no tendrá defectos en los materiales ni en la mano de obra en condiciones normales de utilización y mantenimiento. El periodo de garantía es de tres años y comienza en la fecha de despacho. Las piezas de repuesto, reparaciones y servicios son garantizados por 90 días.
GARANTIA LIMITADA E LIMITAÇÃO DE RESPONSABILIDADE Todos os produtos da Fluke são garantidos contra defeitos de material ou fabricação, sob circunstâncias normais de uso e manutenção. O período de garantia é de três anos, a partir da data da remessa. As peças, reparos e serviços são garantidos por 90 dias.
BEGRÄNSAD GARANTI OCH ANSVARSBEGRÄNSNING Varje Flukeprodukt garanteras vara fri från felaktigheter i material och utförande vid normal användning och service. Garantiperioden är tre år och räknas från leveransdagen. För delar, produktreparationer och service gäller 90 dagars garanti.
有限担保和有限责任 Fluke 担保在正常使用和保养的情况下,其产品没有材料和工艺上的缺陷。从寄 送之日起,担保期为三年。部件、产品修理和服务的担保期限为 90 天。本担保 仅限于 Fluke 授权零售商的原购买人或最终用户,并且不适用于一次性电池、电 缆接头、电缆绝缘转换接头或 Fluke 认为由于误用、改装、疏忽、污染及意外或 异常操作或处理引起的任何产品损坏。Fluke 担保软件能依照功能规格正常运行 90 天,并且软件是记录在无缺陷的媒介上。Fluke 并不担保软件毫无错误或在运 行中不会中断。 Fluke 授权的零售商应仅对最终用户就新的和未使用的产品提供本担保,但无权 代表Fluke 公司提供额外或不同的担保。只有通过 Fluke 授权的销售店购买的产 品或者买方已经按适用的国际价格付款才能享受 Fluke 的担保支持。在一国购买 的产品需在他国修理时,Fluke 有权向买方要求负担重大修理/零件更换费用。 Fluke 的担保为有限责任,由 Fluke 决定是否退还购买金额、免费修理或更换在 担保期间退还 Fluke 授权服务中心的故障产品。 如需要保修服务,请与您就近的 Fluke 授权服务中心
ОГРАНИЧЕННАЯ ГАРАНТИЯ И ОГРАНИЧЕНИЕ ОТВЕТСТВЕННОСТИ Для каждого продукта Fluke гарантируется отсутствие дефектов материалов и изготовления при нормальном использовании и обслуживании. Гарантийный срок составляет три года и отсчитывается от даты поставки. На запчасти, ремонт оборудования и услуги предоставляется гарантия 90 дней.
SINIRLI GARANTİ VE SORUMLULUK SINIRI Tüm Fluke ürünleri, normal kullanım ve servis koşulları altında madde ve işçilik kusurları olmayacağı konusunda garanti altına alınmıştır. Garanti süresi iki yıl olup, ürünün gönderildiği tarihte başlar. Parçalar, ürün onarımları ve servisler, 90 gün için garanti altına alınmıştır.
OGRANICZONA GWARANCJA I OGRANICZENIE ODPOWIEDZIALNOŚCI Każdy produkt firmy Fluke posiada gwarancje na brak usterek materiałowych i produkcyjnych w warunkach normalnego użytkowania i konserwacji. Okres gwarancji obejmuje trzy lata i rozpoczyna się w dniu wysłania produktu. Części, naprawy produktu oraz serwisowanie są objęte gwarancją przez 90 dni.
OMEZENÁ ZÁRUKA A OMEZENÍ ZODPOVĚDNOSTI Firma Fluke garantuje, že každý její výrobek je prost vad materiálu a zpracování při normálním použití a servisu. Záruční doba je tří roky a začíná datem expedice. Díly, opravy produktů a servis jsou garantovány 90 dní.
Table of Contents Title Introduction........................................................................ To Contact Fluke ............................................................... Safety Information ............................................................. Symbols ............................................................................. Unpack the Tester ............................................................. The Tester ...............................................................
1550C/1555 Users Manual Electrical Specifications ................................................. 30 Principle of Measurement and Resistance .......................
List of Tables Table 1. 2. 3. Title Page Symbols ............................................................................. 5 List of Replacable Parts..................................................... 28 Accessories ....................................................................... 29 List of Figures Figure 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. Title Standard Items Provided ................................................... 1550C/1555 Insulation Tester ...................................
1550C/1555 Users Manual iv
Introduction The Fluke 1550C and 1555 Insulation Testers (hereafter “the Tester”) are highvoltage insulation testers to validate general circuits, such as switchgear, motors, and cables.
1550C/1555 Users Manual To Contact Fluke To contact Fluke, call one of the following telephone numbers: • Technical Support USA: 1-800-44-FLUKE (1-800-443-5853) • Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-5853) • Canada: 1-800-36-FLUKE (1-800-363-5853) • Europe: +31 402-675-200 • Japan: +81-3-3434-0181 • Singapore: +65-738-5655 • Anywhere in the world: +1-425-446-5500 Or, visit Fluke's website at www.fluke.com. To register your product, visit http://register.fluke.com.
Insulation Tester Safety Information Safety Information Warning: Read before Tester use. To prevent possible electrical shock or personal injury, follow these guidelines: • Before and after testing, confirm that the Tester does not indicate the presence of a hazardous voltage, see Figure 3. If the Tester beeps continuously and a hazardous voltage is shown on the display, remove power from the circuit under test or allow the installation capacitance to fully discharge.
1550C/1555 Users Manual • • • • • • • • • • 4 Do not exceed the voltage or measurement category (CAT) rating of the test probes/accessories. Not all accessories provided are for use at the full-rated output voltage of the Tester. Accessories rated to 1000V CAT III/ 600V CAT IV are intended for hands-free use during insulation testing and are not to be touched while the output of the Tester exceeds the accessory’s marked rating.
Insulation Tester Symbols Symbols Symbols on the Tester and in the manual are explained Table 1. Table 1. Symbols Symbol Meaning P ® Conforms to European Union standards. ) Canadian Standards Association is the certified body used for testing compliance to safety standards. W Risk of Danger. Important information. See Manual. X Hazardous voltage T Equipment protected by double or reinforced insulation. Examined and licensed by TÜV Product Services.
1550C/1555 Users Manual Unpack the Tester The Tester comes with the items shown in Figure 1. If the Tester is damaged or an item is missing, immediately contact the place of purchase. 1 2 3 5 4 6 7 9 10 8 11 12 GHH01.
Insulation Tester The Tester The Tester The subsequent sections tell you about the Tester and its operation. The Tester is shown in Figure 2. 3 4 2 1 5 7 6 ghh16.eps Item Description Item Description LCD IR Port Safety Shutter Pushbuttons AC plug Built in Handle Input Terminals Figure 2.
1550C/1555 Users Manual Pushbuttons Use the pushbuttons to control the Tester, view test results, and scroll through chosen test results. Pushbuttons and their functionality are discussed in Figure 2. FUNCTION 1 2 3 4 5 ENTER TEST 6 7 Ghh02.eps Item Description Turns the Tester off and on. Push to go to the Function menu. Push again to exit the Function menu. To scroll within the Function menu, use the arrow pushbuttons.
Insulation Tester The Tester Additionally, and are also used to access the following menu items: 1.X Insulation Functions: 1.1 Ramp off (default) 1.2 Ramp on 1.3 DAR T= 01-00 1.4 DAR/PI T= 10-00 2 Time limit xx-xx 3 Show results 4 Delete results Push to make the selection. Turn the Tester On and Off Push to turn on the Tester. The Tester does a self-check, self-calibration, shows the current software version, and starts in the Test Voltage mode.
1550C/1555 Users Manual Display Display annunciators are shown in Figure 4. 5 6 7 4 8 3 9 2 1 10 11 12 GHH03f.EPS Item Description Interference present. Readings could be out of specified accuracy range. Polarization Index. Electrical breakdown in Ramp mode. Dialectric Absorption Ratio. Ramp mode indicator. Possible hazardous voltage is at the test terminals. XW Warning: Before and after testing, confirm that the Tester does not indicate the presence of hazardous voltage.
Insulation Tester The Tester Charge the Battery Pb = Note This Tester uses a rechargeable 12 V lead-acid battery for power. Do not mix with the solid waste stream. Use a qualified recycler or hazardous materials handler to discard dead batteries. Contact your approved Fluke Service Center for disposal and recycling data. Storing rechargeable lead-acid batteries in a low-charged state could decrease their life and/or damage them.
1550C/1555 Users Manual To charge the battery with the ac power supply: 1. Turn the Tester off. 2. Disconnect the test leads from the Tester. 3. Move the safety shutter to access the power supply connection. 4. Connect the ac power cord to the IEC ac power socket on the Tester. 5. Connect the other end of the power cord to an ac power supply. See “General Specifications” for ac charger input specifications. The LCD displays CHARGING. Downloading is possible when the Tester is in Charging mode.
Insulation Tester The Tester Shield ( ) ( ) Surface Leakage Current GHH13.EPS Figure 6. Surface Leakage Current Figure 7 shows how to prevent surface current leakage with a lead connected from the Guard terminal to a conductor that surrounds the inner insulation. The surface leakage current is directed to the Guard terminal. This removes the leakage current from the measurement path between the positive and negative terminals, and improves the accuracy of the test readings.
1550C/1555 Users Manual Shield ( ) (G ) ( ) Leakage Current GHH15.EPS Figure 8. Improved Guard Terminal Connection Measurements Common measurement procedures are discussed in this section. Connect to the Circuit Under Test Warning To avoid possible electric shock or personal injury: • Remove all power from the circuit under test and discharge circuit capacitance before testing a circuit with the Tester.
Insulation Tester Measurements To connect to the circuit under test: 1. Move the safety shutter to access the input terminals. 2. Put the test leads into the correct terminals shown, see Figure 9. 3. Connect the test leads to the circuit under test. 1 3 2 GHH09.EPS (-) Negative Terminal (+) Positive Terminal Safety Shutter Figure 9. Test Lead Connections Note The Tester is NOT specified below 200 kΩ.
1550C/1555 Users Manual Before an Insulation Test The Tester includes features and functions that let you adapt the test to your requirements. These features let you: • define a test voltage • make a ramp test selection • set a time limit (duration) for the test • measure polarization index (PI) • measure dielectric absorption ratio (DAR) • measure capacitance Use these alone or in combination. Set, clear, or account for (as appropriate) each feature before you start an insulation test.
Insulation Tester Measurements Note The actual test voltage can be up to 10 % higher than the selected test voltage. Program a Test Voltage To set a test voltage in between the preset test voltages, proceed as follows: 1. With the Tester turned on, push to select TEST VOLTAGE. Test Voltage Display GHH05.EPS 2. Push or to scroll through the preset test voltage options (250 V, 500 V, 1000 V, 2500 V, 5000 V, and 10,000 V). Select the voltage closest to the level required. 3.
1550C/1555 Users Manual Select a Ramp or Steady-State Test The ramp-test function is an automated test that checks insulation for a breakdown. During a ramp test, the output voltage starts at 0 V and increases linearly (100 V/s) until it reaches the specified test voltage or until a sudden drop in measured resistance is detected. Then, the Ramp stops, the test voltage drops to zero, and the voltage at the breakdown point is stored in memory on the Tester.
Insulation Tester Measurements 3. Push to call the menu item. 4. Push or to select the time. 5. Push or to use the settings. starts the test. Polarization Index (PI) As part of the insulation test, the Tester measures and stores polarization index (PI), when appropriate. A polarization index test requires 10 minutes to complete. Therefore, the Tester will start a countdown at 10 minutes. When an insulation test is 10 minutes or more, the polarization test is completed and stored.
1550C/1555 Users Manual Insulation Test XW Warning To avoid possible electric shock or personal injury: • Be aware that measuring insulation resistance requires the application of potentially dangerous voltages to the circuit. This may include exposed bonded metalwork. • Remove all power from the circuit under test and discharge circuit capacitance before testing a circuit with the Tester. • Before proceeding, ensure that the installation is wired correctly and no personnel are endangered by any tests.
Insulation Tester Measurements XW Warning Before and after testing, confirm that the Tester does not indicate the presence of a hazardous voltage, see Figure 4. If the Tester beeps continuously and a hazardous voltage is shown on the display, disconnect test leads and remove power from the circuit under test. 3. Push for 1 second to start the insulation test.
1550C/1555 Users Manual If breakdown occurs with ramp test enabled, push before going to step 4. Following termination of an insulation test, the Tester beeps when a potentially hazardous voltage remains on the test terminals due to charged-circuit capacitance or from the presence of an external voltage. 4. When the test is terminated, STORE RESULT? Is displayed. If appropriate, store the test results as described in the next procedure. Otherwise, terminate the STORE RESULT? Prompt by pushing .
Insulation Tester Measurements View Results Stored in Memory Note Parameters not appropriate for a test are shown as INVALID. The Tester can store 99 sets of test data, including: • Tags • Ramp on or off • Insulation Resistance • Timer reading at termination of test (Timer) • Test Voltage Selected (TV) • Actual Test voltage (V) • Capacitance I • Polarization Index (PI) • Dielectric absorption ratio (DAR) • Test current (I) • Reason for ending the test • Limit – off or timer setting (1 to 99 minutes) (T.
1550C/1555 Users Manual 1 3 2 GHH06.EPS Item Description Voltage at terminals Stored locations Stored test results Figure 11. Viewing Stored Test Data 4. Push or to step through the stored locations. 5. Stop at the location you want to view. 6. Push to view the stored test data for a specific test. Test data appears on the alphanumeric text display and on the LCD. 7. Push to call the menu selection.
Insulation Tester Measurements 1 2 GHH04.EPS Item Description IR Port IR Device Figure 12. IR Port on 1550C/1555 Insulation Tester Install FlukeView Forms Basic Software FlukeView Forms Basic software must be installed on your PC to download stored test data. To install FlukeView Forms Basic, see the FlukeView Forms Installation Guide. To install the infrared adapter, see the USB-IR Cable Installation Guide.
1550C/1555 Users Manual Download Results to PC Note Before the USB-IR cable can be used, software drivers must be installed on your Windows PC. See the USB-IR Installation Guide for more information. To connect the Tester to the PC for use with FlukeView Forms Basic Documenting Software: The Tester should not be running a test; otherwise, serial communications will be disabled. 1. Connect the USB-IR cable to an available USB port on the PC. 2. Attach the IR device to the IR port on the Tester. 3.
Insulation Tester Maintenance Notes Individual test locations cannot be deleted; however, they can be overwritten. The Delete function deletes all stored test results. Maintenance Warning To avoid possible electric shock or personal injury: • Do not attempt to repair or service your Tester beyond what is described in this manual. • Only qualified service personnel should service the Tester. • There are no user-replaceable parts inside the Tester.
1550C/1555 Users Manual Replaceable Parts and Accessories Table 2 lists the replaceable parts that are available for the Tester. Table 3 identifies the accessories available for use with the Tester. Table 2. List of Replaceable Parts Parts Part No.
Insulation Tester Specifications Table 3. Accessories Accessories Part No. Extended Test Lead Set, 25 feet (7.6 meters) 2032761 Rugged Alligator Clamps 3611951 Soft Case 3592805 Hard Case 3671624 Specifications General Specifications Display 75 mm x 105 mm Power 12 V lead-acid rechargeable battery. 2.6 Ahr Charger Input (AC) 85 V to 250 V ac, 50/60 Hz, 20 VA This Class II (double insulated) instrument is supplied with a Class 1 (grounded) power cord.
1550C/1555 Users Manual Electromagnetic Compatibility EN 61326-1, EN 61326-2-2 Certifications P ® ) Safety Compliance EN 61010-1, EN 61557 Parts 1 and 2 CAT III 1000V, CAT IV 600V Pollution Degree 2 Typical Battery Charge Capability Test Voltages Note At temperature extremes, the battery needs to be charged more frequently. Number of Tests 250 V 4100 500 V 3600 1 kV 3200 2.
Insulation Tester Specifications 1000 V 2500 V 5000 V 10,000 V < 200 kΩ unspecified 200 kΩ to 20 GΩ 5% 20 GΩ to 200 GΩ 20 % > 200 GΩ unspecified < 200 kΩ unspecified 200 kΩ to 50 GΩ 5% 50 GΩ to 500 GΩ 20 % > 500 GΩ unspecified < 200 kΩ unspecified 200 kΩ to 100 GΩ 5% 100 GΩ to 1 TΩ 20 % > 1 TΩ unspecified < 200 kΩ unspecified 200 kΩ to 200 GΩ 5% 200 GΩ to 2 TΩ 20 % > 2 TΩ unspecified Bar graph range: Insulation test voltage accuracy: Induced ac mains current rejection: Ch
1550C/1555 Users Manual Short circuit current > 1 mA and < 2 mA Principle of Measurement and Resistance The Tester measures insulation parameters and displays the results using with the following formulas.