Data Sheet
2 Fluke Corporation Fluke 1555 and 1550C Insulation Resistance Testers
1550C 5 kV Insulation Tester
1555 10 kV Insulation Tester
1550C/Kit 5 kV Insulation Tester Kit
1555/Kit 10 kV Insulation Tester Kit
Optional accessories
TL1550EXT 25 foot extended
test lead set
TLK1550-RTLCT Heavy Duty Alligator
Clip leads
Included accessories
Test Cables with Alligator Clips
(red, black, green)
Infrared adapter with interface cable
FlukeView Forms Basic CD-ROM
AC Power Cord
Soft Carrying Case (base models only)
English Manual
Users Manual on CD-ROM
Quick Reference Card
Software License Agreement
Registration Card
FlukeView Forms Installation Guide
USB-IR Cable Installation Guide
IP67 Hard Case (kit only)
Certificate of Calibration (kit only)
Heavy Duty Alligator Clip Leads
(kit and 1555 only)
Software specifications
Fluke ViewForms basic software
requires a PC running Windows XP,
Windows Vista , Windows 7 and
Windows 8.
Electrical specifications
The tester’s accuracy is specified for one year after calibration at operating
temperatures of 0 °C to 35 °C. For operating temperatures outside the range
(-20 °C to 0 °C and 35°C to 50 °C), add ± .25 % per °C, except on the 20 % bands
add ± 1 % per °C.
Insulation resistance measurement
Test voltage (dc) Range Accuracy (± reading)
250 V < 250 kΩ
250 kΩ to 5 GΩ
5 GΩ to 50 GΩ
> 50 GΩ
unspecified
5 %
20 %
unspecified
500 V < 500 kΩ
500 kΩ to 10 GΩ
10 GΩ to 100 GΩ
> 100 GΩ
unspecified
5 %
20 %
unspecified
1000 V < 1 MΩ
1 MΩ to 20 GΩ
20 GΩ to 200 GΩ
> 200 GΩ
unspecified
5 %
20 %
unspecified
2500 V < 2.5 MΩ
2.5 MΩ to 50 GΩ
50 GΩ to 500 GΩ
> 500 GΩ
unspecified
5 %
20 %
unspecified
5000 V < 5 MΩ
5 MΩ to 100 GΩ
100 GΩ to 1 TΩ
> 1 TΩ
unspecified
5 %
20 %
unspecified
10000 V (1555 Only) < 10 MΩ
10 MΩ to 200 GΩ
200 GΩ to 2 TΩ
> 2 TΩ
unspecified
5 %
20 %
unspecified
Bar graph range 0 to 1 TΩ (1550C) - 0 to 2 TΩ (1555)
Insulation test voltage accuracy -0 %, +10 % at 1 mA load current
Induced ac mains current rejection 2 mA maximum
Charging rate for capacitive load 5 seconds per μF
Discharge rate for capacitive load 1.5 s/μF
Range Accuracy
Leakage current
measurement
1 nA to 2 mA ± (5 % + 2 nA)
Capacitance measurement 0.01 uF to 15.00 μF ± (15 % rdg + 0.03 μF)
Timer Range Resolution
0 to 99 minutes Setting: 1 minute
Indication: 1 second
Live circuit warning Warning range Voltage accuracy
30 V to 660 V ac/dc, 50/60 Hz ± (15 % + 2 V)
General specifications
Display 75 mm x 105 mm
Power 12 V lead-acid rechargeable battery, Yuasa NP2.8-12
Charger Input (AC) 85 V to 250 V ac 50/60 Hz 20 VA
This Class II (double insulated) instrument is supplied with a Class 1 (grounded) power cord. The protective earth terminal
(ground pin) is not connected internally. The extra pin is for added plug retention only.
Dimensions 269 mm x 277 mm x 160 mm (10.6 in x 10.9 in x 6.3 in)
Weight 3.7 kg (8.2 lb)
Temperature (operating) -20 ºC to 50 ºC (-4 ºF to 122 ºF)
Temperature (storage) -20 ºC to 65 ºC (-4 ºF to 149 ºF)
Humidity 80 % to 31 ºC decreasing linearly to 50% at 50 ºC
Altitude 2000 m
Enclosure sealing IP40
Input overload protection 1000 V ac
Electromagnetic compatibility EN 61326
Certifications
Safety compliance IEC 61010-1: Overvoltage categroy II, pollution degree 2.
IEC 61010-2-030: Measurement 600V CAT IIV / 1000V CAT III
IEC 61557-1, IEC 61557-2
Fluke Corporation
PO Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2005-2014 Fluke Corporation.
Specifications subject to change without notice.
Printed in U.S.A. 10/2014 6002295B_EN
Modification of this document is not permitted
without written permission from Fluke Corporation.
Fluke. Keeping your world
up and running.
®
Ordering information
(RF connection time (binding
time) can take up to 1 minute.)