Thermometer Readout User's Guide
Table Of Contents
- Table of Contents
- 1 Before You Start 1
- 2 Introduction 7
- 3 Specifications and Environmental Conditions 9
- 4 Quick Start 13
- 5 Parts and Controls 17
- 6 General Operation 21
- 7 Menu Functions 29
- 7.1 Channel Menu 29
- 7.2 Probe Menu 35
- 7.2.1 Edit Probe 36
- 7.2.1.1 ITS-90 Conversion 38
- 7.2.1.2 ITS-SR5 39
- 7.2.1.3 PT-100 Conversion 40
- 7.2.1.4 Callendar-Van Dusen (CVD) conversion 40
- 7.2.1.5 RES Conversion 41
- 7.2.1.6 Thermistor T(R) Conversion [THERM-T] 41
- 7.2.1.7 Thermistor R(T) Conversion [THERM-R] 42
- 7.2.1.8 YSI-400 Conversion 42
- 7.2.1.9 Thermocouple Volts [Vin[mV]] 42
- 7.2.1.10 Standard Thermocouple Conversions 43
- 7.2.1.11 Thermocouple Polynomial Conversion 43
- 7.2.2 Copy Probe 44
- 7.2.3 Test Probe 44
- 7.2.4 Print Probe 45
- 7.2.5 Default Probe 46
- 7.2.1 Edit Probe 36
- 7.3 Fields Menu 47
- 7.4 Logging Menu 53
- 7.5 System Menu 69
- 8 Digital Communications Interface 79
- 8.1 Overview 79
- 8.2 Communications 79
- 8.3 Interface Commands 80
- 8.4 Commands 86
- 8.4.1 Measurement Commands 88
- 8.4.2 Measurement Control Commands 91
- 8.4.2.1 INITiate 91
- 8.4.2.2 INITiate:CONTinuous? 91
- 8.4.2.3 SENSe:AVERage:COUNt? [MIN|MAX|DEF] 91
- 8.4.2.4 SENSe:AVERage:COUNt
|MIN|MAX|DEF 91 - 8.4.2.5 SENSe
:RESistance:RANGe? [MIN|MAX|DEF] 91 - 8.4.2.6 SENSe
:RESistance:RANGe |MIN|MAX|DEF 92 - 8.4.2.7 SENSe
:RESistance:WIRE? [MIN|MAX|DEF] 92 - 8.4.2.8 SENSe
:RESistance:WIRE |MIN|MAX|DEF 93 - 8.4.2.9 TRIGger:TIMer? [MIN|MAX|DEF] 93
- 8.4.2.10 TRIGger:TIMer
|MIN|MAX|DEF 93
- 8.4.3 Channel Commands 94
- 8.4.4 Probe Commands 95
- 8.4.4.1 CALCulate
:CONVert:CATalog? 95 - 8.4.4.2 CALCulate
:CONVert:COPY |ALL 96 - 8.4.4.3 CALCulate
:CONVert:NAMe? 96 - 8.4.4.4 CALCulate
:CONVert:NAMe 96 - 8.4.4.5 CALCulate
:CONVert:PARameter:CATalog? 97 - 8.4.4.6 CALCulate
:CONVert:PARameter:VALue? [|ALL] 97 - 8.4.4.7 CALCulate
:CONVert:PARameter:VALue , [,< param>, ...] 97 - 8.4.4.8 CALCulate:CONVert:PRINt [
|ALL [, ]] 98 - 8.4.4.9 CALCulate
:CONVert:SNUMber? 98 - 8.4.4.10 CALCulate
:CONVert:SNUMber 98 - 8.4.4.11 CALCulate
:CONVert:TEST? | 99
- 8.4.4.1 CALCulate
- 8.4.5 Calibration Coefficient Commands 99
- 8.4.5.1 CALibrate:AUTo 99
- 8.4.5.2 CALibrate
:DATE:CALibrate? [MIN|MAX|DEF] 99 - 8.4.5.3 CALibrate
:DATE:CALibrate ( , , )|MIN|MAX|DEF 100 - 8.4.5.4 CALibrate
:DATE:DUE? [MIN|MAX|DEF] 100 - 8.4.5.5 CALibrate
:DATE:DUE ( , , )|MIN|MAX|DEF 100 - 8.4.5.6 CALibrate
:PARameter:OFFSet ? [MIN|MAX|DEF] 101 - 8.4.5.7 CALibrate
:PARameter:OFFSet |MIN|MAX|DEF 101 - 8.4.5.8 CALibrate
:PARameter:SCALe ? [MIN|MAX|DEF] 101 - 8.4.5.9 CALibrate
:PARameter:SCALe |MIN|MAX|DEF 102 - 8.4.5.10 CALibrate
:PARameter:LINearity ? [MIN|MAX|DEF] 102 - 8.4.5.11 CALibrate
: PARameter:LINearity |MIN|MAX|DEF 103 - 8.4.5.12 CALibrate
:PARameter:RJC? [MIN|MAX|DEF] 103 - 8.4.5.13 CALibrate
:PARameter:RJC |MIN|MAX|DEF 103
- 8.4.6 Display Commands 104
- 8.4.6.1 DISPlay:RESolution? [MIN|MAX|DEF] 104
- 8.4.6.2 DISPlay:RESolution
|AUT|MIN|MAX|DEF 104 - 8.4.6.3 DISPlay:LAMP? [MIN|MAX|DEF] 104
- 8.4.6.4 DISPlay:LAMP
| |MIN|MAX|DEF 104 - 8.4.6.5 DISPlay:DECimal:FORMat? [MIN|MAX|DEF] 105
- 8.4.6.6 DISPlay:DECimal:FORMat
|MIN|MAX|DEF 105 - 8.4.6.7 DISPlay:WINDow? [MIN|MAX|DEF] 105
- 8.4.6.8 DISPlay:WINDow
|MIN|MAX|DEF 105 - 8.4.6.9 DISPlay:WINDow
:FIELd :FEED? 105 - 8.4.6.10 DISPlay:WINDow
:FIELd :FEED [, ] 106
- 8.4.7 Logging Commands 106
- 8.4.7.1 LOGging:AUTomatic:DELete [
|ALL] 106 - 8.4.7.2 LOGging:AUTomatic:COUNt? [MIN|MAX|DEF] 106
- 8.4.7.3 LOGging:AUTomatic:COUNt
|MIN|MAX|DEF 106 - 8.4.7.4 LOGging:AUTomatic:FREE? 107
- 8.4.7.5 LOGging:AUTomatic:LABel? [MIN|MAX|DEF] 107
- 8.4.7.6 LOGging:AUTomatic:LABel
|MIN|MAX|DEF 107 - 8.4.7.7 LOGging:AUTomatic:POINt? [MAX] 107
- 8.4.7.8 LOGging:AUTomatic:PRINt [
|ALL [, ]] 108 - 8.4.7.9 LOGging:AUTomatic:STATus? 108
- 8.4.7.10 LOGging:AUTomatic:STATus
108 - 8.4.7.11 LOGging:AUTomatic:TIMe? [MIN|MAX|DEF] 108
- 8.4.7.12 LOGging:AUTomatic:TIMe
|MIN|MAX|DEF 109 - 8.4.7.13 LOGging:AUTomatic:VALue?
|MIN|MAX|DEF 109 - 8.4.7.14 LOGging:DEMand:DELete [
|ALL] 109 - 8.4.7.15 LOGging:DEMand:FREE? 109
- 8.4.7.16 LOGging:DEMand:LABel? [MIN|MAX|DEF] 110
- 8.4.7.17 LOGging:DEMand:LABel
|MIN|MAX|DEF 110 - 8.4.7.18 LOGging:DEMand:POINt? 110
- 8.4.7.19 LOGging:DEMand:PRINt [
|ALL [, ]] 110 - 8.4.7.20 LOGging:DEMand:STORe 111
- 8.4.7.21 LOGging:DEMand:VALue?
|MIN|MAX|DEF 111 - 8.4.7.22 LOGging:LABel
:NAME? 111 - 8.4.7.23 LOGging:LABel
:NAME
- 8.4.7.1 LOGging:AUTomatic:DELete [
- 8.4.8 System Commands 111
- 8.4.9 Communication Interface Commands 114
- 8.4.9.1 SYSTem:COMMunicate:IR:FEED? [MIN|MAX|DEF] 114
- 8.4.9.2 SYSTem:COMMunicate:IR:FEED
|MIN|MAX|DEF 114 - 8.4.9.3 SYSTem:COMMunicate:IR:MODE? [MIN|MAX|DEF] 114
- 8.4.9.4 SYSTem:COMMunicate:IR:MODE
|MIN|MAX|DEF 114 - 8.4.9.5 SYSTem:COMMunicate:IR:TIMe? [MIN|MAX|DEF] 115
- 8.4.9.6 SYSTem:COMMunicate:IR:TIMe
|MIN|MAX|DEF 115 - 8.4.9.7 SYSTem:COMMunicate:SERial:BAUD? [MIN|MAX|DEF] 115
- 8.4.9.8 SYSTem:COMMunicate:SERial:BAUD
|MIN|MAX|DEF 115 - 8.4.9.9 SYSTem:COMMunicate:SERial:FDUPlex? [MIN|MAX|DEF] 115
- 8.4.9.10 SYSTem:COMMunicate:SERial:FDUPlex
|MIN|MAX|DEF 116 - 8.4.9.11 SYSTem:COMMunicate:SERial:FEED? [MIN|MAX|DEF] 116
- 8.4.9.12 SYSTem:COMMunicate:SERial:FEED
|MIN|MAX|DEF 116 - 8.4.9.13 SYSTem:COMMunicate:SERial:LINefeed? [MIN|MAX|DEF] 116
- 8.4.9.14 SYSTem:COMMunicate:SERial:LINefeed
|MIN|MAX|DEF 117 - 8.4.9.15 SYSTem:COMMunicate:SERial:TIMe? [MIN|MAX|DEF] 117
- 8.4.9.16 SYSTem:COMMunicate:SERial:TIMe
MIN|MAX|DEF 117 - 8.4.9.17 SYSTem:KLOCkout? [MIN|MAX|DEF] 117
- 8.4.9.18 SYSTem:KLOCkout
|MIN|MAX|DEF 117 - 8.4.9.19 SYSTem:POWer:BATTery? 118
- 8.4.9.20 SYSTem:POWer:SOURce? 118
- 8.4.10 Date and Time Commands 118
- 8.4.10.1 DISPlay:DATE:FORMat? [MIN|MAX|DEF] 118
- 8.4.10.2 DISPlay:DATE:FORMat
|MIN|MAX|DEF 118 - 8.4.10.3 DISPlay:TIME:FORMat? [MIN|MAX|DEF] 119
- 8.4.10.4 DISPlay:TIME:FORMat
|MIN|MAX|DEF 119 - 8.4.10.5 SYSTem:DATE? [MIN|MAX|DEF] 119
- 8.4.10.6 SYSTem:DATE
, , 119 - 8.4.10.7 SYSTem:TIME? 120
- 8.4.10.8 SYSTem:TIME
, , 120
- 8.4.11 Password Commands 120
- 8.4.12 Status Commands 122
- 8.4.12.1 *CLS 122
- 8.4.12.2 *ESE? 122
- 8.4.12.3 *ESE
|MIN|MAX|DEF 122 - 8.4.12.4 *ESR? 122
- 8.4.12.5 *SRE? 122
- 8.4.12.6 *SRE
|MIN|MAX|DEF 122 - 8.4.12.7 *STB? 122
- 8.4.12.8 *TST? 123
- 8.4.12.9 STATus:MEASure? 123
- 8.4.12.10 STATus:MEASure:CONDition? 123
- 8.4.12.11 STATus:MEASure:ENABle? 123
- 8.4.12.12 STATus:MEASure:ENABle
123 - 8.4.12.13 STATus:OPERation? 124
- 8.4.12.14 STATus:OPERation:CONDition? 124
- 8.4.12.15 STATus:OPERation:ENABle?[MIN|MAX|DEF] 124
- 8.4.12.16 STATus:OPERation:ENABle
|MIN|MAX|DEF 124 - 8.4.12.17 STATus:QUEStionable? 124
- 8.4.12.18 STATus:QUEStionable:CONDition? 125
- 8.4.12.19 STATus:QUEStionable:ENABle? [MIN|MAX|DEF] 125
- 8.4.12.20 STATus:QUEStionable:ENABle
|MIN|MAX|DEF 125
- 8.4.13 Statistical Calculation Types 126
- 8.4.14 Field Types 126
- 8.4.15 Conversion Types and Parameters 128
- 8.4.16 Port Numbers 129
- 8.4.17 Date and Time Formats 130
- 9 Calibration 131
- 10 Maintenance 137
- 11 Troubleshooting 139
- Figures
- Figure 1 Using the Clamp-on Ferrites 14
- Figure 2 Front Panel 17
- Figure 3 Back Panel 18
- Figure 4 Thermocouple Connections 23
- Figure 5 Probe Connection Wiring Diagram 24
- Figure 6 Main Menu 29
- Figure 7 Measure Period 30
- Figure 8 Channel Menu 30
- Figure 9 Enable Channel 32
- Figure 10 Channel Mode 33
- Figure 11 Moving Average 34
- Figure 12 Display Options 35
- Figure 13 Probe Menu 36
- Figure 14 Edit Probe 37
- Figure 15 Copy Probe 44
- Figure 16 Test Probe 45
- Figure 17 Print Probe 46
- Figure 18 Default Probe 47
- Figure 19 Clear Stats 48
- Figure 20 Fields Menu 48
- Figure 21 Select Fields 49
- Figure 22 Edit Fields 51
- Figure 23 Default Fields 52
- Figure 24 Display Options 53
- Figure 25 Logging Menu 54
- Figure 26 Demand Log Submenu 55
- Figure 27 Store Readings 56
- Figure 28 Log History 57
- Figure 29 Demand Log View Data 58
- Figure 30 Print Data 59
- Figure 31 Delete Data 60
- Figure 32 Logging Options 61
- Figure 33 Auto Log Submenu 61
- Figure 34 Log Statistics (Auto Log Started) 63
- Figure 35 Start Stop 63
- Figure 36 Auto Log View Data 64
- Figure 37 Print Data 65
- Figure 38 Delete Data 66
- Figure 39 Log Stats 67
- Figure 40 Data Labels 68
- Figure 41 Default Labels 69
- Figure 42 System Menu 70
- Figure 43 Comm Setup 71
- Figure 44 Date Time 73
- Figure 45 Password 74
- Figure 46 Password for Calibration Access 75
- Figure 47 Select Calibration Channel 76
- Figure 48 Pass Cal Reset 77
- Figure 49 Serial Cable Wiring 79
- Figure 50 Using a Shorting Wire 132
- Tables
- Table1 International Electrical Symbols 1
- Table 2 Conversion Types 38
- Table 3 Alphabetical List of Commands 81
- Table 4 Statistical Calculation Types 126
- Table 5 Field Types 126
- Table 6 Reference (REF) Keywords 127
- Table 7 Conversion Types, Mnemonics, and Serial Port Responses 128
- Table 8 Conversion Types and Parameters 128
- Table 9 RJC Settings 129
- Table 10 Port Numbers 129
- Table 11 Date Formats 130
- Table 12 Time Formats 130
- Table 13 PRT and Thermistor Calibration Parameter Description 131
- Table 14 Calibration Parameters 135
- Index
- !
- A
- B
- C
- D
- E
- F
- G
- I
- L
- M
- P
- R
- S
- T
- U
- V
- W
- Y

3.
Connect a 400Ω resistor (6 ppm accuracy) to the input and measure its
resistance. Note the average error in the measurement. Adjust the C400
parameter by subtracting the measured error. For example, if the input is
exactly 400.0000Ω and the readout shows 399.9913Ω, adjust the C400
parameter by adding 0.0087.
4. Record the date in the DATE parameter and set the DUE parameter to
the desired calibration interval.
5.
Verify the accuracy at 0Ω, 25Ω or 50Ω, 100Ω, 200Ω, and 400Ω on each
channel.
9.1.3 Thermistor Calibration Parameters
Three adjustable parameters are used for calibration: C0K, C10K, and C100K.
The C0K directly affects the measurement at 0Ω. It has negligible effect at 10
kΩ but significant affect at higher resistances. The C10K directly affects the
measurement at 10 kΩ. It has negligible effect at 0Ω and proportionately
greater effect the higher the resistance. The C100K directly affects the mea-
surement at 100 kΩ. It has negligible effect at 0Ω and10kΩ with greater effect
as the resistance increases. Each of the parameters has positive effect at their
primary resistances: increasing the value of the parameter increases the mea-
sured resistance. The theoretically normal value for each is 0.
The calibration parameters C0K, C10K, and C100K can be adjusted to opti-
mize the accuracy. These parameters are accessed from the CALIBRATION
function of SYSTEM menu.
9.1.4 Thermistor Calibration Procedure
Calibration requires adjustment of the C0K, C10K, and C100K parameters at
three specific input resistances. If the resistances used are approximately 0Ω,
10 kΩ and 100 kΩ respectively the adjustments are independent and the proce
-
dure is simple. The order in which the adjustments are performed is important.
The adjustment of the C100K parameter must be done last as the adjustments
of C0K and C10K affect the measurement at 100 kΩ but C100K does not affect
the measurements at 0Ω or 10 kΩ. Each channel must be calibrated. Set the
conversion type to RES to display resistance and set the RANGE parameter to
10KΩ. The accuracy required of the resistance standards is 1/4 of the instru
-
ment accuracy; that is 0.12Ω at 0Ω,0.25Ω (25 ppm) at 10 kΩ, and 2.5Ω (25
ppm) at 100 kΩ. The recommended procedure is as follows:
1.
Connect a 0Ω resistor to the input and measure its resistance. If a short
-
ing wire is used the wire should run from the C2 terminal to the C1 ter
-
minal to the P1 terminal to the P2 terminal (See Figure 50 on page 132).
Note the average error in the measurement. Adjust the C0K parameter by
subtracting the measured error. For example, if the input is exactly 0.0Ω
and the readout shows –0.11Ω, adjust the C0K parameter by adding
0.11.
133
9 Calibration
PRT and Thermistor Calibration