Data Sheet

Cursor measurements (125, 124)
Sources: Input A, Input B
Modes: Single or dual vertical cursor, dual horizontal
cursor, risetime or falltime
Measurements:
Single vertical line: Average, min value, max value,
time from start of recording in roll mode
Dual vertical lines: V at markers, time between
cursors, 1/T between cursors (in Hz)
Dual horizontal lines: High, low or V – readout,
risetime and falltime: transition time, 0 %-level,
100 %-level, with markers at 10 % and 90 %
Accuracy: As oscilloscope
Bus health tester (Fluke 125 only)
Bus health automatically analyzes the electrical signals
on the network to give waveform data and measure
individual parameters. Automatic comparison of the
measurement results to the standards results in “good”
or “false” indicators to be displayed per parameter.
Bus types and reference standards used:
AS-i (EN50295, 166 kb/s)
®
;
CAN-bus (ISO-11898, up to 1 Mb/s);
Interbus S (EIA-485, up to 10 Mb/s)
®
;
ControlNet (61158 type 2, 2.5 Mb/s)
®
;
Modbus (EIA-232, up to 115 kb/s and
EIA-485, up to 10 Mb/s)
®
;
Foundation Fieldbus H1 (61158 type 1, 31.25 kb/s)
and H2 (61158 type 1, up to 10 Mb/s);
Profibus DP (EIA-485, up to 10 Mb/s) and
PA (61158 type 1, 31.25 kb/s);
Ethernet [10Base2 (coaxial) and 10BaseT (UTP)], 10 Mb/s;
RS-232 (EIA-232, up to 115 kb/s);
RS-485 (EIA-485, up to 10 Mb/s);
or user defined system
Measured parameters: Baud rate, risetime, falltime,
high level, low level, distortion, amplitude and jitter,
with comparison to system’s standard values
Power measurements (Fluke 125 only)
Measure types: Watt, VA, VAR, Power Factor (PF)
Power configuration: Single phase or balanced
3-phase (delta-configuration) mains supply
Voltage measurement: Channel A, using STL120,
voltage probe or direct input
Current measurement: Channel B, using i400s
current clamp (included) or other compatible clamp
Current clamp or shunt sensitivity:
0.1/1/10/100/1000 mV/A, 10 mV/mA and 400 mV/A
Harmonics mode (Fluke 125 only)
Converts waveform information into a harmonics dis-
play (using FFT processing), which shows the relative
amplitudes of the 1st up to the 33rd harmonic
Harmonics frequency range: DC to 33rd harmonic
(fundamental 60 Hz); DC to 24th (fundamental
400 Hz)
Analyzed waveform: Voltage waveform (Ch.A),
current waveform (Ch.B) or power (Ch.A x Ch.B),
automatically generated
Display: Bargraph showing 1st up to 33rd harmonic,
amplitude displayed in % relative to fundamental
Timebase setting: 5 ms/div.
Measurements: Relative amplitude of individual
harmonic; THD in %r or %f
Oscilloscope mode
Vertical deflection
Bandwidth and risetime Fluke 125, 124 Fluke 123
Bandwidth (risetime)
with VPS40 probes
input A and B directly
with STL120 Shielded
Test Leads
40 MHz
40 MHz
12.5 MHz
20 MHz
20 MHz
12.5 MHz
Instrument risetime
(input directly)
8.75 ns 17.5 ns
Number of inputs: two
Input coupling: AC, dc with ground level indicator
Input sensitivity: 5 mV to 500 V/div (with the
included VPS40 (Fluke 125, 124) and STL120 shielded
test leads measure up to 600 Vrms, CAT III)
Input voltage: 600 V CAT III. See “general specifica-
tions” for more detailed information.
Vertical resolution: 8 bit
Accuracy: ± (1 % of reading + 0.05 x range/div)
Input impedance: 1 MW ± 1 % // 225 pF with
STL120 shielded test leads; 1 MW ± 1 % // 20 pF
± 3 pF with BB120; 5 MW ± 1 % // 15.5 pF with
VPS40, 10:1 voltage probe
Horizontal
Max. sample rate (both channels simultaneously):
Fluke 125, 124: 2.5 GS/s for repetitive signals;
25 MS/s for single shot
Fluke 123: 1.25 GS/s for repetitive signals; 25 MS/s
for single shot
Number of digitizers: two
Time base range: 10 ns/div to 1 min/div
(Fluke 125, 124)
Maximum record length: 512 Min/Max points
per input
Accuracy: ± (0.1 % of reading + 1 pixel)
Glitch detect: 40 ns
Display and acquisition
Display modes: Input A, input A and B, envelope,
smooth
Acquisition modes: Normal (including glitch capture),
single shot, roll
Trigger and delay
Source: Input A, input B, external via optional ITP120
Modes: Automatic Connect-and-View, Free Run, Edge,
Single Shot, Video, Video Line
Connect-and-View: Advanced automatic triggering
that recognizes signal patterns and automatically sets
up and continuously adjusts triggering, time base and
amplitude. Automatically displays stable pictures of
complex and dynamic signals like motor drive and
control signals
Video triggering: NTSC, PAL, PAL+, SECAM. Includes
line select
Time delay: Up to 10 divisions pre-trigger view
Measurements
Vdc, Vac, Vac+dc, Vpeak max, Vpeak min, Vpeak
to peak, frequency (Hz), positive pulse width, negative
pulse width, positive duty cycle, negative duty
cycle, Amp ac, Amp dc, Amp ac+dc, Phase,
Temperature °C, Temperature °F, dBV, dBm into 50 W
and 600 W. (Amps, °C or °F with optional probes)
ScopeMeter 215C, 225C and 190C Series, 120 Series Fluke Corporation 5
Technical specifications 120 Series