Data Sheet
ScopeMeter
®
120 and 190 Series
incl. 225C and 215C
Technical Data
ScopeMeter 215C,
225C and 190C Series:
Speed, performance
and analysis power
For demanding applications, the
ScopeMeter 215C, 225C and 190C
Series high-performance oscillo-
scopes offer specifications usually
found on top-end bench instru-
ments. With up to 200 MHz band-
width, 2.5 GS/s real time sampling
and a deep memory of 27,500
points per input, they’re ideal
for engineers who need the full
capabilities of a high-performance
scope in a handheld, battery pow-
ered instrument.
•
Dual input - 200, 100 or
60 MHz bandwidth
•
Up to 2.5 GS/s real time
sampling per input
•
Bus Health Test capability for
industrial buses (225C and 215C)
•
High waveform resolution of
3000 datapoints per channel
•
Frequency Spectrum using FFT
analysis
•
Connect-and-View
™
automatic
triggering, a full range of manual
trigger modes plus external trig-
gering
•
Digital Persistence for analyzing
complex, dynamic signals like on
an analog oscilloscope
•
Fast display update rate for
seeing dynamic behavior
instantaneously
•
Automatic capture and replay of
100 screens
•
27,500 points and more per
input record length using
ScopeRecord
™
mode
•
1,000 V CAT II and 600 V
CAT III safety certified
•
Up to 1,000 V independently
floating isolated inputs
•
5000 count DMM and paperless
recorder built-in
•
Four hours rechargeable
NiMH battery pack
ScopeMeter 120 Series:
Three-in-one simplicity
The compact ScopeMeter 120 Series
is the rugged solution for industrial
troubleshooting and installation
applications. It’s a truly integrated
test tool, with oscilloscope, multi-
meter and “paperless” recorder in
one affordable, easy-to-use instru-
ment. Find fast answers to problems
in machinery, instrumentation,
control and power systems.
•
A dual input 40 MHz or 20 MHz
digital oscilloscope
•
Two 5,000 counts true-rms
digital multimeters
•
Cursor measurements
(Fluke 124, 125)
•
A dual input TrendPlot recorder
•
Connect-and-View trigger
simplicity for hands-off operation
•
Shielded test leads for oscillo-
scope, resistance, continuity
and capacitance measurements
•
Full bandwidth, VPS40 10:1
40 MHz probe included standard
with Fluke 124, 125
•
Up to seven hours battery
operation
•
600 V CAT III safety certified
•
Optically isolated RS-232
interface
•
Rugged, compact case
•
Bus Health test for industrial bus
systems (Fluke 125)
•
Power measurements and har-
monics measurement (Fluke 125)