Application Note

4 Fluke Corporation Using a Fluke ScopeMeter 125 to Troubleshoot FOUNDATION™ Fieldbus Installations
user can enter other limits at set
up limits on the front menu.
For Fieldbus systems, selecting
9 volts as the minimal value and
32 volts as the maximum value
will generally do the job.
During use, the Fluke 125 will
indicate via icons whether the
measured voltage is within the
limits: = OK; X = not-OK. In
addition, the icon may change
to a warning sign (!) when a
reading is within a certain per-
centage of a limit value.
Figure 5, a screen shot, shows
an actual bus-health test screen
for an H1 Fieldbus system. If the
instrument recognizes that com-
munication is taking place, the
activity indicators will blink.
The first line displays bias
voltage. OK indicates that the
dc-supply voltage (27.7 volts) is
within the limit values: 9.0 and
32.0 volts.
The Fluke 125 differs from a
standard DMM in that it displays
the voltage reading that is clos-
est to a limit over the time span
that the measurement is happen-
ing. The instrument has a hold
function that helps installation
and maintenance personnel
determine if the supply voltage
is at risk of going outside of the
preset limits. In other words,
during load changes, the instru-
ment displays the value closest
to the lower or upper limit.
To reset the hold function,
press the Hold/Run key twice
to hold and restart the mea-
surements. This action initiates
a new measurement cycle
in which all result fields are
cleared.
The instrument can be used as
a standard DMM or as a standard
oscilloscope. Then, it can record
momentary voltages in order to
isolate instabilities in a power
supply. In addition, the instru-
ment can record supply-voltage
changes on the bus. The function
that allows this is TrendPlot
,
detailed in the instruments
users manual.
The hour-glass icon on the
fourth line of Figure 5 indicates
that a rise-time measurement
was in progress at the moment
the screen-copy was made. Next
to the icon is the result of an
earlier measurement and the
limit to which this rise time is
compared. In this application,
rise times up to 8 µs are viewed
as acceptable.
Because of inevitable voltage
drop based on copper resistance
and Ohm’s-law, supply voltage to
devices near the end of a trunk
are lower that for those nearer
the power supply. Making bias
voltage measurements at various
junctions can reveal disruptions,
such as poor connections, along
the trunk. Good notes and knowl-
edge of the trunk layout will help
troubleshooters in finding which
junction/spur is at fault.
In H1 Fieldbus systems, the
maximum cable length in the
trunk is 1900m. For a twisted
pair cable made from AWG#18
(1 mm in diameter or with a
cross section of 0.79 mm
2
per
conductor), one must take into
account a resistance of 2.26
Ω per 100m of wire, which, of
course, translates to 4.52 Ω per
100 m of dual-conductor cable
and a total resistance of 86 Ω for
the two wires over the maximum
trunk length.
If there is only a single device
drawing 25 mA connected to the
end of the trunk, that device in
itself will cause a voltage drop of
2.2 volts across the trunk cable.
With multiple devices connected
along the line, the voltage drop
due to supply-current consump-
tion can make the difference
between a good and a poor sup-
ply voltage at some devices.
Table 1 (following Figure 6)
shows the calculated supply
voltages for the network in
Figure 6, which depicts a full-
length trunk with a limited
number of devices. In the design
phase of a new network, similar
calculations should be made to
determine the type of cabling
and the power supply needed.
For existing systems, if a
plant’s archives include “as-
builtswith design data and
information about the sys-
tem’s physical layout and cable
Figure 5: Fluke 125 screen showing results of a Bus Health Test.
Figure 6: Calculation of voltage drops resulting from connected devices and their distances from the power supply.
25 mA50 mA75 mA100 mA
25 mA
1900 m
4
25 mA
1000 m500 m100 m0 m
0 V 0.452 V 1.81 V 2.94 V 3.96 V
Distance
DC Power
Supply
Voltage
Drop
Cable: AWG 18 per conductor (= 1 mm diameter)
Resistance: 2.26 ohm per 100 m per conductor
3
25 mA
2
25 mA
1