Datasheet
Technical Specifications
CableIQ™ Qualification Tester and SimpliFiber® Pro
Fluke Corporation
P.O. Box 777, Everett, WA USA 98206-0777
Fluke Networks operates in more than 50 countries
worldwide. To find your local office contact details,
go to www.flukenetworks.com/contact.
©
2009 Fluke Corporation. All rights reserved.
PrintedinU.S.A.5/20093474891D-ENG-NRevA
NETWORKSUPERVISION
M12/ RJ45 Cable Specifications
Cable type Ethernet cable, Cat5e, shielded, 2 Pair
AWG 26 stranded (7 wire), RAL 5021
(water blue), M12 4 pos. D- coded on
RJ45 connector
Number of positions 4
Fixed cable length 2 m
Volume resistance ≤ 5 mΩ
Insulation resistance ≥ 100 MΩ
Ambient temperature -20
o
C to 50
o
C
Inflammability class acc
to UL 94
V0
Surge voltage category II
Pollution degree 3
Degree of protection IP20/IP67
External cable diameter 6.7 mm
Transmission
characteristics
Cat 5 (IEC 11801:2002), Cat 5e (TIA
568B:2001)
CableIQ™ Specifications (continued)
SimpliFiber® Pro Specifications
SimpliFiber® Pro Specifications (continued)
General Specifications
Temperature range Operating: -10
0
C to 50
0
C
Storage: -20
0
C to 50
0
C
Humidity range 95% (10
0
C to 35
0
C) non-condensing;
75% (35
0
C to 40
0
C) non-condensing;
uncontrolled <10
0
C
Certifications CE, CSA, N10140, Class 1 laser-safe
Dimensions Power meter: 6.4 in x 3.2 in x 1.5 in
(16.5 cm x 8.0 cm x 3.9 cm)
MM/SMsources:5.6inx3.2inx1.6in
(14.2 cm x 8.1 cm x 4.1 cm)
Weight Power meter: 11.5 oz (325 g)
MM/SMsources:9.8oz(278g)
Optical Sources
Optical output connector Fixed SC
Emitter type 850/1300:LED
1310/1550:FPLaser
FindFiber: Laser
Emitter wavelengths 850, 1300, 1310, 1490, 1550, 1625
Power output (minimum) MM: ≥ -20 dBm
SM: ≥ 8 dBm minimum;
-7 dBm nominal
Power output stability
(8 hours)
MM:+/-0.1dBover8hours
SM:+/-0.25dBover8hours
MM battery life
(2 x AA IEC LR6)
40 hours typical
SM battery life
(2 x AA IEC LR6)
30 hours typical
FindFiber battery life
(2 x AA IEC LR6)
80 hours typical
Optical Power Meter
Power measurement
accuracy
+/-0.25dB
Optical connector Removable adapter; SC adapter standard;
Optional adapters include LC, ST
Detector type InGaAs
Calibrated wavelengths 850, 1300, 1310, 1490, 1550, 1625
Power measurement range 850: 10 to -52 dBm
1300, 1310, 1490, 1550, 1625: 10 to -60
dBm
Power measurement
linearity
850nm:+/-0.2dB;+/-0.2dBforpower
from0dBmto-45dBm,+/-0.25dBfor
power < -45 dBm; 1300 nm, 1310 nm,
1490nm,1550nm,1625nm:+/-0.1dB;
+/-0.1dBforpowerfrom0dBmto-55
dBm,+/-0.2dBforpower>0dBmand<
-55 dBm
Resolution 0.01 dB
Battery life >50hourstypical
Memory 1000 loss or power measurements
Serial communication
physical interface
USB