Datasheet

Maximum number of
incidents shown
9
Distance accuracy
(0 m to 1500 m or
0 ft to 4921 ft)
± (1 m + 0.1 % x length) for reflective incidents
1
± (3 m + 0.1 % x length) for non-reflective
incidents
2
Testing speed
< 6 seconds typical
Connector
Removable/cleanable SC adapter, UPC polish
Fiber types tested
50/125 µm or 62.5/125 µm multimode
Detection of
reflective incidents
3
-35 dB default threshold (User selectable: -20 dB to
-45 dB in 5 dB increments)
Reflectance
accuracy
4
± 4 dB
Maximum reflectance
measurement
-20 dB
Detection of loss
incidents
5
0.70 dB default threshold (user-configurable from
0.5 dB to 6.1 dB in 0.2 dB increments)
Bulkhead quality
If no fiber is attached or if the connector is dirty,
the troubleshooter displays 0 m or 0 ft.
Live fiber detection
Detects optical signals from 600 nm to 1050 nm
and shows ACTIVE LINE if a signal is there. Looks
for a signal every 3 seconds after the first detec-
tion. +7 dB maximum input power.
Certifications and
compliance
P Conforms to relevant European Union directives
; Conforms to relevant Australian standards
$ Listed by the Canadian Standards Association
CSA C22.2 No. 61010.1.04
Conforms to FCC Rules, Part A, Class A
1. ± user-configurable Index of Refraction (IOR) error ± the incident location error. Incident location error for reflective incidents: ±1 m from 1 m to 1.5 km.
2. ± user-configurable Index of Refraction (IOR) error ± the incident location error. Incident location error for non-reflective incidents: ±2 m for lengths 15 m,
otherwise ±1 m.
3. Finds and gives the location of an incident that has a reflectance larger than -55 dB. Detects incidents >1m after the bulkhead connector when the
bulkhead reflectance is <35 dB. Detects incidents >3 m after an incident when the incident reflectance is <35 dB.
4. With a backscatter coefficient of -63 dB at 850 nm using a calibrated -14 dB reference.
5. Detects incidents >10 m after the bulkhead connector or any prior incident when the bulkhead reflectance is <-35 dB and the reflectance of any prior
incident is <-35 dB. The maximum link loss prior to the incident is <7 dB.
Specifications
Operating temperature
with the battery
0ºC to 50ºC
Non-operating
temperature
-20ºC to 60ºC
Operating relative
humidity
(without condensation)
95% (10ºC to 35ºC)
75% (35ºC to 40ºC)
uncontrolled < 10ºC
Vibration
Random, 5 Hz to 500 Hz, MIL-PRF-28800F CLASS 2
Shock
1 meter drop test
Altitude
3000 m
EMC
EN 61326-1: 2004
Battery type
2 AA alkaline batteries (no battery charger)
Battery life
1500 tests (typical)
Laser safety
Class 1 CDRH
Complies to EN 60825-2
LCD type
Backlit black and white (segments)
Index of refraction
range
1.45 to 1.5 (factory default is 1.496)
Auto turn off
Automatically turns off after 5 minutes if no
keys are pressed. Backlight turns off first.
Factory calibration
interval
None
Output wavelengths
850 nm ± 10 nm
Laser classification
Class 1 CDRH Complies to EN 60825-2
Dynamic range
>11 dB
Maximum distance
1500 meters or 4921 feet