Instructions

Table Of Contents
SimpliFiber Pro Meter and Test Kits
Users Manual
24
To set the reference:
1
Clean the connectors on the meter, source, and a test
reference cord. Use fiber optic solvent and optical-
grade wipes or swabs to clean connectors as described
on page 13.
2
Turn on the meter and source and let them warm up
for 5 minutes. Allow additional time if the equipment
has been stored above or below ambient temperature.
3
Make the connections shown Figure 10.
4
If the source’s AUTO LED is not on, press .
If you want to set the reference for both wavelengths,
press
until the wavelength LEDs blink alternately.
Note
If you are not using a SimpliFiber Pro source, set the
source to the desired wavelength and continuous
wave output.
5
On the meter, press until POWER appears (see
Figure 8 on page 21).
Notes
You cannot set the reference if Min or Max appear
on
the display. See page 22.
Setting the reference from power mode lets you
see the s
ource’s power level before saving it as the
reference level. You may also set the reference
from loss mode.
6
Press SET REF. The meter switches to loss mode, the
display shows  dB, OK appears briefly, and the new
reference value appears.
If the reference value is less than -60 dBm or if the
source is
in ID or 2 kHz mode, the meter briefly shows
and . Check the connections and the source’s
mode or try setting the reference again using another
test
reference cord.
Caution
If you disconnect the source’s output after setting
the reference, you must set the reference again to
ensure valid measurements.