Operation Manual

Custom Test Settings
Creating a Custom Outlet Configuration
12-3
12
To create a custom twisted pair limit:
1
Turn the rotary switch to SETUP; then select Twi sted
Pair.
2
Select Test Limit, press J More; then select Custom.
3
Press J Create.
4
On the Custom screen, select Name; then use the text
editing screen to enter a name for your custom limit.
Press
N when you are done.
5
On the Custom screen, select Use Default Values From,
select a test limit group; then select a limit as a baseline
for your custom limit.
6
The Custom screen shows other settings on multiple
tabs for the twisted pair limit. Use AD and
H to select settings to change. Use BC to move
among the tabs.
See page 3-3 for details on twisted pair test settings.
7
From the Custom screen, press N when you are done
creating the limit. The tester saves the custom limit in
the Custom list, with asterisks (*) before and after the
name.
Creating a Custom Outlet
Configuration
You can create up to nine custom outlet configurations. A
custom outlet configuration includes the following settings:
Custom outlet name
Baseline outlet configuration for default settings
Testing enabled or disabled for each wire pair
To create a custom outlet configuration:
1
Turn the rotary switch to SETUP; then select Twi ste d
Pair.
2
Select Outlet Configuration; then select Custom.
3
Press J Create.
4
On the Custom screen, select Name; then use the text
editing screen to enter a name for your custom
configuration. Press
N when you are done.
5
On the Custom screen, select Use Default Values From;
then select an outlet configuration as a baseline for
your custom configuration.
-continued-