Operation Manual

DTX Series CableAnalyzer
Technical Reference Handbook
6-4
Table 6-1. Diagnosing Twisted Pair Test Failures (cont.)
Characteristic impedance exceeds the limit or an anomaly is detected
Bad connection
Cable compression (tight cable ties, pinches, kinks, etc.)
Mismatch of cable types
Water in cable jacket
Excessive loading at coaxial cable tap
Incorrect terminator value (coaxial cable)
Length gives a FAIL result
Cable is too long (may need to remove coiled service loops)
NVP is set incorrectly
Length exceeds the limit, but the length test did not fail
If the length of the shortest pair does not exceed the limit by 10 %, then the length test passes even if other pairs exceed the
limit. This is the 10 % rule for length, as given in the ANSI/TIA standard. See the Fluke Networks Knowledge base for details.
Propagation delay or delay skew gives FAIL result
Cable is too long (may need to remove coiled service loops)
Cable uses different insulation materials on different pairs
Insertion loss gives FAIL, FAIL*, or PASS* result
Cabling is too long
Poor quality patch cord
Bad connection
Wrong cable type in installation
Wrong test standard selected