Operation Manual
DTX Series CableAnalyzer
Technical Reference Handbook
6-4
Table 6-1. Diagnosing Twisted Pair Test Failures (cont.)
Characteristic impedance exceeds the limit or an anomaly is detected
•
Bad connection
•
Cable compression (tight cable ties, pinches, kinks, etc.)
•
Mismatch of cable types
•
Water in cable jacket
•
Excessive loading at coaxial cable tap
•
Incorrect terminator value (coaxial cable)
Length gives a FAIL result
•
Cable is too long (may need to remove coiled service loops)
•
NVP is set incorrectly
Length exceeds the limit, but the length test did not fail
If the length of the shortest pair does not exceed the limit by 10 %, then the length test passes even if other pairs exceed the
limit. This is the 10 % rule for length, as given in the ANSI/TIA standard. See the Fluke Networks Knowledge base for details.
Propagation delay or delay skew gives FAIL result
•
Cable is too long (may need to remove coiled service loops)
•
Cable uses different insulation materials on different pairs
Insertion loss gives FAIL, FAIL*, or PASS* result
•
Cabling is too long
•
Poor quality patch cord
•
Bad connection
•
Wrong cable type in installation
•
Wrong test standard selected