Twisted Pair Test Settings
Table 2 describes the settings that apply to twisted pair
cabling tests.
To access the settings, turn the rotary switch to SETUP, use
to highlight Twisted Pair; then press .
Table 2. Twisted Pair Test Settings
Setting Description
Select the appropriate test limit for the job. Selecting Cu
stom lets you create a test limit. See the
Technical Reference Handbook for details.
Select a cable type appropriate for the type you will test. The cable types are organized by type and
manufacturer. Selecting Custom lets you create a cable type. See the Technical Reference Handbook
for details.
Nominal velocity of propagation, which is used with the measured propagation delay to determine
cable length. The default value defined by the selected cable type represents the typical NVP for that
cable type. You may enter a different value if necessary. To determine the actual value, change the
NVP until the measured length matches the known length of a cable. Use a cable at least 30 m
(100 ft) long.
Increasing the NVP increases measured length.
The Outlet Config
uration setting determines which cable pairs are tested and which pair numbers
are assigned to the pairs. See Figure 10. To see the wire map for a configuration, press
Sample
from the Outl
et Configuration screen. Selecting “Custom” lets you create a configuration. See the
Technical Reference Handbook for details.
-continued-
SETUP > Twisted
Pair > Test Limit
SETUP > Twisted
Pair > Cable
Typ e
SETUP > Twisted
Pair > NVP
SETUP > Twisted
Pair > Out
let
Configuration