User's Manual

TiS10, TiS20, TiS40, TiS45, TiS50, TiS55, TiS60, TiS65
Users Manual
24
To set the IR-Fusion
mode:
1. Push 2.
2. Push W/X to highlight Image.
3. Push 1 or Z to view the menu.
4. Push W/X to highlight IR-Fusion.
5. Push 1 or Z to view the menu.
6. Push W/X to highlight an option.
7. Push:
1 to set the change and go back to the live
view.
2 or Y to set the change and go back to the
previous menu.
3 to cancel the change and go back to the live
view.
Color Alarms
The Imager has apparent temperature color alarms. The
type of available alarm depends on the model.
Model: Hi-Lo Alarm: Isotherm Alarm:
TiS60, TiS65, TiS50, TiS55 yes yes
TiS40, TiS45 yes not available
TiS10, TiS20 not available not available
The high-temperature color alarm shows a full visible
image and only shows infrared information on objects or
areas that are above the set apparent temperature alarm
level. The low-temperature (or dew point) color alarm
shows a full visible image and only shows infrared
information on objects or areas that are below the set
apparent temperature (or set dew point) color alarm level.
The user must manually determine and set these
parameters.
Note
The Imager does not sense ambient or surface
dew point level automatically. To use the low-
temperature color alarm function as a dew point
color alarm, manual determination and input of
surface dew point temperature will yield the best
results. Depending on the situation, the colors
presented may help identify areas of concern with
possible dew point condensation.
To view the Color Alarm menu:
1. Push 2.
2. Push W/X to highlight Image.
3. Push 1 or Z to view the menu.
4. Push W/X to highlight Color Alarm.
5. Push 1 or Z to view the menu.