3000 FC
Users Manual
38
AC and DC Current
Function
Range
[1]
Resolution
Accuracy
(45 Hz to 1 kHz)
60.00 mA 0.01 mA
1.5 % + 3
400.0 mA
[3]
0.1 mA
[2]
60.00 mA 0.01 mA
0.5 % + 3
400.0 mA
[3]
0.1 mA
[1] All ac current ranges are specified from 5 % of range to 100 % of range.
[2] Input burden voltage (typical): 400 mA input 2 mV/mA.
[3] 400.0 mA accuracy specified up to 600 mA overload.
Frequency
Range Resolution
Accuracy
[1]
99.99 Hz 0.01 Hz
0.1 % + 1
999.9 Hz 0.1 Hz
9.999 kHz 0.001 kHz
99.99 kHz 0.01 kHz
[1] Frequency is specified up to 99.99 kHz in volts and up to 10 kHz in amps.