To replace the batteries:
1. Turn off the Product and remove all test leads.
2. Lift the tilt stand up as shown in Figure 11.
3. Turn the battery-door latch until the unlock symbol
() aligns with the arrow.
4. Lift off the battery door.
5. Remove the three AA batteries and replace them
with new ones. Use the correct battery orientation.
6. Install the battery door.
7. Turn the battery-door latch until the locked symbol
() aligns with the arrow.
To replace the fuse:
1. Turn off the Product and remove all test leads.
2. As shown in Figure 11, remove six screws from the
case bottom.
3. Pull the case bottom from the case top.
4. Remove the fuse from its holder and replace it with a
440 mA, 1000 V FAST fuse with a minimum interrupt
rating of 10,000 A. Use only Fluke PN 943121.
To reassemble the Product, do the steps in the opposite
sequence.