Wireless Multimeter
Features
13
Table 4. Rotary Switch Positions
Switch
Position
Function
DC voltage from 1 mV to 1000 V.
Push
to measure frequency from 2 Hz to 99.99 kHz
AC voltage measurement from 60.0 mV to 1000 V.
Push to measure frequency from 2 Hz to 99.99 kHz.
Push
again to measure Volts/Hertz.
DC voltage measurements from 1 mV to 600 mV.
Push to measure ac voltage from 6 mV to 600 mV.
[1]
Resistance measurements from 0.1 Ω to 50 MΩ.
Push
to measure capacitance from 1 nF to 9999 μF.
Continuity. Beeper turns on at <25 Ω and turns off at >250 Ω.
Push
for diode test. Shows OL above 2.0 V.
AC current measurements from 3.00 mA to 400 mA.
Push to measure dc current from 3.00 mA to 400 mA.
[1]
Push
again to measure frequency from 2 Hz to 9.99 kHz.
[1] This function will stay in ac or dc when the function switch is moved to another position and back to this function. Even when turned to off and back to
this function.