User manual
Fluke DAQ
Users Manual
The dialog box entries include:
CH #
Computed channels are assigned channel numbers from
901 to 960. 901 to 960 in the 2680A and X501 to X520
where X represents the input module (1-3), on the 2638A
and 1586A.
Label
Use the label field to identify where the computed channel
is used.
Unit Label
The value to apply to the label. For example, the
calculated value can be shown as Watts.
Function
• Take the average reading from one channel.
• Subtract the measurement of one channel from the
measurement of another channel.
• Generate an equation based on measurements.
Alarm Type
Use Normal alarm if the measurement goes above or
below the Alarm 1 or Alarm 2 level. Use Band to alarm if
the signal falls between two levels (2680A and NetDAQ).
Alarm 1 or Alarm 2
Each channel has two alarms. The alarms can be set to
Off, Low, or High. An alarm condition occurs when a
measurement falls below a low alarm value (Low) or rises
above a high alarm value (High). The Software records all
alarm conditions in a data file.
Digital Output
Used to output a digital signal or close a relay if an alarm
is detected.
Use Channel as Alarm Trig.
Select if channel reading will be used to trigger an alarm.
Trigger Scan With Module
Calculate the computed channel any time a module is
scanned (2680A and NetDAQ).
Channel A or Channel B
Select from the configured channels in the device.
Note
Channels in another device cannot be selected.
32