Factors (Mx + B)
Used to scale measurements. Mx + B scaling multiplies a
measurement by a multiplier (M) and then applies an
offset (B). For example, Mx + B scaling of 100x + 50
applied to a measured value of 1.15 would result in a
reading of 100(1.15) + 50 = 165.
Input or Scaled Range
Use the M and B values to calculate the Input and Scaled
range or the Input and Scaled range can be entered and
the Software will calculate the M and B values.
Configure a Computed Channel
Computed channels provide a means to calculate values
based on measurements. For example, if Channel 1
measures dc V and Channel 2 measures dc I, a
computed channel could be used to calculate
instantaneous power (Channel 1 x Channel 2).
For a computed channel to execute, it must be associated
with the scanning of a module or instrument. Choose a
module in the Scan with Module area for this association.
The default is module 1.
To configure a computed channel:
1. Highlight the Computed Channels icon in the
TreeView panel show and click Insert.
2. Specify the number of channels to insert on the
Insert Channel dialog box shown below.
alg117s.bmp
3. Click OK. The computed channels are added and are
shown in the TreeView panel.
4. Select a computed channel. The Computed Channel
settings dialog is shown.