User manual

Product Overview and Specifications
Measurement Specifications 1
1-19
Resistance Input Characteristics
Range Resolution
Resolution
Source Current
Fast
4½ Digits
Medium
5 ½ Digits
Slow
6 ½ Digits
100 100.0000 10 m 1 m 0.1 m
1 mA / 4 V
1 k 1.000000 k 100 m 10 m 1 m
1 mA / 4 V
10 k 10.00000 k 1 100 m 10 m
100 µA / 6 V
100 k 100.0000 k 10 1 100 m
100 µA / 12 V
1 M 1.000000 M 100 10 1
10 µA / 12 V
10 M 10.00000 M 1 k 100 10
1 µA / 12 V
100 M 100.0000 M 10 k 1 k 100
0.1 µA / 12 V
Resistance Accuracy
Accuracy is given as ± (% measurement + % of range). Basic accuracy specification is for 4-wire resistance. For 2-wire
resistance add 0.02 internal resistance if using Channel 1 or 1.5 if using channels x01 through x20, and add external
lead wire resistance.
Range
24 Hour
(23 ±1 °C)
90 Days
(23 ±5 °C)
1 Year
(23 ±5 °C)
T.C./ °C
Outside 18 °C
to 28 °C
100
0.003 % + 0.003 % 0.008 % + 0.004 % 0.01 % + 0.004 % 0.0006 % + 0.0005 %
1 k
0.002 % + 0.0005 % 0.008 % + 0.001 % 0.01 % + 0.001 % 0.0006 % + 0.0001 %
10 k
0.002 % + 0.0005 % 0.008 % + 0.001 % 0.01 % + 0.001 % 0.0006 % + 0.0001 %
100 k
0.002 % + 0.0005 % 0.008 % + 0.001 % 0.01 % + 0.001 % 0.0006 % + 0.0001 %
1 M
0.002 % + 0.001 % 0.008 % + 0.001 % 0.01 % + 0.001 % 0.001 % + 0.0002 %
10 M
0.015 % + 0.001 % 0.02 % + 0.001 % 0.04 % + 0.001 % 0.003 % + 0.0004 %
100 M
0.3 % + 0.01 % 0.8 % + 0.01 % 0.8 % + 0.01 % 0.05 % + 0.002 %
Note:
For conducted disturbances on mains input >1 V from 10 MHz to 40 MHz, add 0.6 % of range. For disturbances >3 V,
accuracy is unspecified.
Resistance Additional Errors
Digits NPLC Additional NPLC Noise Error
6 ½ 200 0 % of range
6 ½ 100 0 % of range
6 ½ 10 (Slow) 0 % of range
5 ½ 1 (Medium) 0.0008 % of range
4 ½ 0.2 (Fast)(only for 2-wire)
0.002 % of range + 12 m
4 ½ 0.02 (only for 2-wire)
0.01 % of range + 17 m
RTD
Temperature Range .............................................. -200 °C to 1200 °C (depending on the sensor)
Resistance Range ................................................. 0 to 4 k
Maximum Lead Resistance (4-wire ) ................ 2.5 % of range per lead for 400 and 4 k ranges
Sample Rate
Slow .................................................................... 10 PLC
Medium ............................................................... 2 PLC
Fast ..................................................................... 1 PLC