Error Code Error Message Causes and Solutions
Error 206
Error reading cold
junction temperature
Cause: Failed to read cold junction temperature from module.
Solution: Cycle the power. If the error occurs again, contact
Fluke Calibration (see Chapter 1).
Error 207 Failed to reset relay
on slot 1
Cause: The relay does not reset when the channel is
switched.
Solution: Cycle the power and try again. If the error occurs
again, the relay card needs serviced, contact Fluke
Calibration (see Chapter 1).
Error 208 Failed to reset relay
on slot 2
Error 209 Failed to reset relay
on slot 3
Error 260 Error copying
configure to ChXXX
Cause: The copy function failed.
Solution: Check the channel list and make sure all selected
channel are set up correctly.
Error 300 Invalid procedure
Cause: Wrong calibration procedure is specified in
CALibrate:STARt command or MAIN in CALibrate:STARt
command could be misspelled.
Solution: Check the parameter value and try again.
Error 301
Invalid procedure
step
Cause: Wrong or misspelled calibration step is specified in
CALibrate:STARt command.
Solution: Check the parameter value and try again.
Error 302
Step change not
allowed while busy
Cause: Invalid CALibrate:STARt was received while a
calibrate operation is in progress.
Solution: Send CALibrate:ABORt to terminate the active
calibration procedure.
Error 303 Invalid starting point
Cause: The step specified in CALibrate:STARt command
should be the first point of a function or a range, it is invalid to
start a calibration procedure from an intermediate step.
Solution: Check the start point and try again.
Error 304 Invalid reference unit
Cause: CALibrate:NEXT command was received with an
unexpected unit.
Solution: Check the parameter value and try again.
Error 305
Entered value out of
range
Cause: CALibrate:NEXT command was received, but the
reference value is out of range.
Solution: Check the parameter value and try again.