User manual
  Scan/Monitor, Record, and Data 
  Scan  4 
4-11 
•  For the Alarm trigger type: The scan sweep is triggered by an alarm that has 
tripped. To do this, select the Alarm trigger type in the test setup and assign a 
channel to be a trigger. When the scan is started, the Product automatically starts 
to monitor the channel set as the trigger for a tripped alarm. Once an alarm is 
tripped, the scan sweep starts and runs as directed by the test setup. Scan sweeps 
will continue as long as the alarm is tripped. If the alarm is no longer tripped, the 
Product will finish the scan in progress then wait for next alarm trip. 
a.  Configure the applicable channel alarm as previously described. 
b.  Push 
 then push  to start the scan. 
•  For the 
Manual trigger type: Push  to start the scan. The scan performs one 
scan cycle then stops. Push  again to trigger another scan cycle. At any 
time, push  again to stop the test or push  to pause the test. If automatic 
recording is set to OFF, push  to record data. 
View Scan Data and Statistics 
The Scan Data feature lets the user see the scan data from the last scan sweep (see 
Figure 4-4). This feature is located in the Scan menu (push 
 then push ). The 
data on this menu updates in real-time. From the Scan Data menu, the user can open a 
Statistics menu to view minimum, maximum, average, and standard deviation statistics 
for each channel (see Table 4-4). 
If Mx+B is applied to the channel, an [Mx+B] icon will show next to the channel 
number. If an alarm is tripped, the alarm number will show next to the channel as a red 
icon and also the measurement value will be red. See Figure 4-4. 
hcn055.eps 
Figure 4-4. Scan Data 