User manual
2638A 
Users Manual 
4-6 
hcn032.eps 
Figure 4-3. Test Setup Menu Example 
Trigger Type 
The Trigger Type tells the Product when and how to start and stop a scan. There are four 
trigger types: 
Note 
If a Scan Count of 0 is set or if the scan interval is shorter than the scan 
sweep time, the scan continuously repeats until the scan is stopped or the 
Product runs out of memory. 
•  The 
Interval trigger type sets the scan to happen on a user-defined numeric interval 
that is started by a front-panel softkey in the Scan/Monitor menu. The user sets the 
number of times to scan (Scan Count) and how often the scans occur (Interval). 
•  The External trigger type set the scan to start when a set TRIG input on the Digital 
I/O port detects a low condition. Like Interval, the user manually sets the number of 
times to scan (Scan Count) and the time between the scans (Interval). 
•  The 
Alarm trigger type sets the scan to start when a HI or LO channel alarm is tripped 
on a monitored channel. Like Interval, the user manually sets the number of times to 
scan (Scan Count) and the time between the scans (Interval). 
•  The 
Manual trigger type sets the scan to happen only when the user manually pushes 
the Scan/Monitor key on the front panel. This trigger type lets the user set the number 
of times to scan (Scan Count) when the key is pushed, but does not have an interval 
time because the interval is a manual key push. 