User manual
  Input and Channel Configuration 
  Channel Configuration  3 
3-21 
Thermistor Channels 
Refer to Table 3-8 to configure a thermistor channel. 
Table 3-8. Thermistor Channel Setup 
1
2
3
4
5
6
hce022.eps
Item Function 
  Thermistor function selection. 
  Select the thermistor type. Types available: 2.252 kΩ, 5 kΩ, and 10 kΩ. 
 
Set up the channel for a 2-wire measurement connection. See “Sense Input Configuration” on 
page 3-3. 
  Input a custom alpha-numeric label to help identify the channel (optional). 
  Set an alarm for this channel (optional). See “HI and LO Channel Alarms” on page 3-30. 
  Apply Mx+B scaling to the measurement (optional). See “Mx+B Scaling” on page 3-29. 