User manual
  Input and Channel Configuration 
  Input Wiring  3 
3-5 
Input Types and Wiring Diagrams 
Table 3-1 lists the type of input types the Product can measure and the wiring polarity 
that should be used to wire it to the Input Module. 
Table 3-1. Types of Inputs 
Type of Input  Range and Types 
Channel 
Configuration 
Reference 
Wiring Polarity 
ac and dc 
Voltage 
Range: 0 V to 300 V 
[2]
 Page 3-17  
H
L
V
ac and dc 
Current 
[1]
Range: 0 mA to ±100 
mA 
Page 3-17 
Resistance (Ω) 
2-Wire or 4-Wire 
[3]
Range: 0 Ω to 100 MΩ 
Page 3-18 
H
L
R
2-Wire 
H
Source
Sense
L
R
4-Wire 
H
L
H
Source
Sense
L
R
3-Wire 
H
L
Platinum 
Resistance 
Thermometer 
(PRT) 
2-Wire, 3-Wire or 4-Wire 
Types: 
• PT385 or PT392 
• R0: 0.09 Ω to 1.2 kΩ 
Page 3-22 
Thermistor 
2-Wire or 4-Wire 
Types: 
• 2.252 kΩ 
• 5 kΩ 
• 10 kΩ 
Page 3-20 
Thermocouple 
Types: B, C, D, E, G, J, 
K, L, M, N, R, S, T, U, 
W, 
Page 3-19 
H
L
Note 
[1] − Each Input Module has two terminal sets (mA 21 and mA 22) that are dedicated to current measurements. 
[2] − 300 V is for Ch001 only. Other analog channels are 150 V max. 
[3] − 100 M
Ω is for Ch001 only. Other analog channels are 10 MΩ Max. 










