User manual
  Initial Setup and Configuration 
  Input Module and Relay Card Installation  2 
2-7 
Table 2-2. Instrument Setup Menu (cont.) 
Menu Item  Description  Selections Available 
Screen Saver  Change the wait time or disable the screensaver. 
Never, 15 Min, 30 Min, or 
60 Min 
Calibration 
Date 
Shows the date that the Product was last calibrated. See 
the 2638A Calibration Manual for more information. 
-- 
Password 
Management 
Change the Admin and User profile passwords. 
See “Set Up Security” in this 
chapter for more information. 
Resume Scan 
Sets the Product to automatically turn on and resume 
scanning and recording after a power loss. 
See “Automatic Power Loss 
Scan Resume” in Chapter 4 
for more information. 
Remote Port 
Configure the LAN Ethernet or serial USB communication 
settings. 
See the 2638 Remote 
Programmers Guide for more 
information. 
Relay Counter 
Shows how many times the channel relays have been 
scanned along with the name, serial number, and 
calibration date of the Input Module. 
-- 
Input Module and Relay Card Installation 
Before an Input Module can be inserted into the rear panel, a relay card must be installed. 
Use the procedure below and refer to Figure 2-6 as necessary to install the relay card: 
1.  Power off the Product with the main power switch. 
2.  Remove the six screws that secures the plastic frame to the rear panel. 
3.  Remove the plastic frame. 
4.  Slide the aluminum slot protector out of the Product. 
5.  Carefully align the rails of the relay card into the slot guides. 
6.  Slowly push the relay card into the Product until the card is fully seated. 
WCaution 
Do not force the relay card into the slot. The card should easily 
move when the rails of the relay card are properly aligned in the 
slot guides. 
The 2638A/05 has only one available relay slot. Do not remove 
the cover plates on slots 2 and 3. This will damage the Product 
and require a return for service. 
7.  Install the plastic frame and secure it to the rear panel with six screws. 