P5514, P5514-V & P5514-EP Hydraulic Comparison Test Pump Users Manual PN 3952337 November 2010 © 2010 Fluke Corporation. All rights reserved. Printed in USA. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
LIMITED WARRANTY AND LIMITATION OF LIABILITY Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is one year and begins on the date of shipment. Parts, product repairs, and services are warranted for 90 days.
Table of Contents Title Preparation ......................................................................................................... How to Contact Fluke ........................................................................................ Operating Fluid Compatibility........................................................................... Safety Information ............................................................................................. Compressed Liquid........................
P5514, P5514-V & P5514-EP Users Manual ii
List of Tables Table 1. 2. 3. 4. 5. Title Operating Fluid Compatibility ............................................................................... Symbols.................................................................................................................. Test Port Insert - Parts List..................................................................................... Seal Kit - Part Numbers .........................................................................................
P5514, P5514-V & P5514-EP Users Manual iv
List of Figures Figure Title 1. 2. 3. 4. 5. 6. Connecting the Gauge ............................................................................................ Test Port Insert ....................................................................................................... Reservoir and Test Port Assemblies....................................................................... Screw Press Assembly ...........................................................................................
P5514, P5514-V & P5514-EP Users Manual vi
Preparation The Hydraulic Comparison Test Pump User’s Manual covers the following instruments: the P5514, P5514-V, and the P5514-EP. The Comparison Test Pump is used for checking pressure-measuring instruments against Master Test Gauges. Note The terms ”Master Test Gauge” and “Gauge” in this document refer to any pressure-measuring instrument such as Transfer Standards, Digital Calibrators, and Transducers. This system is only as accurate as the Master Test Gauge used.
P5514, P5514-V & P5514-EP Users Manual Operating Fluid Compatibility The standard system is designed for use with a wide range of fluids; however, the O-ring seals are Nitrile; use of solvents, fuel oils, brake fluids or other, similar aggressive fluids can damage the seals. The table below details the two alternative versions of this instrument, which offer increased compatibility with other fluids. Table 1.
Hydraulic Comparison Test Pump Connections Table 2. Symbols Symbol B J W ~ Description AC (Alternating Current) Earth Ground Important Information: refer to manual Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycling information. Connections Fit the device under test (DUT) to the test port using the method described below: W Caution Ensure that all devices are internally clean and free from contamination before connecting to the tester.
P5514, P5514-V & P5514-EP Users Manual 1 2 3 4 Figure 1. Connecting the Gauge gjq001.eps Connect the Gauge Refer to Figure 1 for steps 1 to 4. 1. Screw the appropriate gauge adapter fully on to the instrument to be tested. 2. Screw assembly down COUNTER-CLOCKWISE on to test port. Note Hand-tight is sufficient; ensure that the bottom face contacts the O-ring on the test port. 3.
Hydraulic Comparison Test Pump Operation 1 2 3 4 gjq002.eps Figure 2. Test Port Insert Table 3. Test Port Insert - Parts List Item Description P5515 P5515-V P5515-EP 3867888 1 O Ring 3865142 3865163 2 Test Port Insert 3919892 — 3 O Ring 3883397 3883521 4 Test Port 3918298 — — 3867895 — To calibrate panel-mounted gauges with pressure connections in the rear, use an Angle Adapter (see Ancillary Equipment section). Operation Priming 1.
P5514, P5514-V & P5514-EP Users Manual Operation W Warning Do not exceed the maximum working pressure of the instrument (stated on the instrument label). 1. Turn the screw press in to generate pressure. 2. Allow a few moments for the system to stabilize before taking any readings, especially after large changes in system pressure.
Hydraulic Comparison Test Pump Maintenance and Servicing 4. The white, anti-extrusion ring (21) is a PTFE spiral, and can be removed by “unwinding” it from the rambler. 5. When removing the rambler seal (20), take care not to use any tool that may have a sharp edge that will scratch the surfaces of the rambler, otherwise it may leak when reassembled. 6. The replacement rambler seal can be eased over the front of the rambler, and into the groove. 7.
P5514, P5514-V & P5514-EP Users Manual W Caution If the rambler shows signs of wear, then it is very likely that the main bore in the barrel is worn also. If the bore is worn or scratched, it will not seal correctly, and may leak under pressure. 11. Re-assembly is the reverse of the above. Ensure that all seals and sealing surfaces are clean and undamaged. 13 12 11 10 9 8 7 6 14 14 4 4 5 3 1 3 1 2 2 15 Figure 3. Reservoir and Test Port Assemblies 8 gjq003.
Hydraulic Comparison Test Pump Maintenance and Servicing 17 18 19 26 16 25 21 20 22 23 24 25 26 Figure 4. Screw Press Assembly gjq004.
P5514, P5514-V & P5514-EP Users Manual Table 5.
Hydraulic Comparison Test Pump Ancillary Equipment Ancillary Equipment Angle Adapter, P5543 To calibrate gauges with the pressure connection on the rear (e.g. panel-mount gauges) in their correct position, an angle adapter should be used. The angle adapter fits directly onto the test station, converting it through 90 degrees, allowing the standard adapters to be used. Figure 5. Angle Adapter gjn022.
P5514, P5514-V & P5514-EP Users Manual 12