2 Fluke Calibration P5500 Series hydraulic and pneumatic comparison test pumps, media separators and accessories
Pneumatic comparison test pumps
Model P5510-2M Model P5513-20M
Pressure range 0 psi to 300 psi (2 MPa) 0 psi to 3,000 psi (21 MPa)
Vacuum range 0 in to 27 in Hg (915 mbar 90% vacuum)
Test port adapters M20 X 1.5, M14 X 1.5, 1/8, 1/4, 3/8, and 1/2 NPT and BSP
(plus 1 additional 1/4 NPT adapter)
M20 X 1.5, M14 X 1.5, 1/8, 1/4, 3/8, and 1/2 NPT and BSP
(plus 1 additional 1/4 NPT adapter)
Weight 3.5 kg (7.5 lb) 8 kg (17 lb)
Dimensions (W x D x H) 240 mm x 330 mm x 180 mm (9.5 in x 13 in x 7 in) 440 mm x 300 mm x 215 mm (17.5 in x 12 in x 8.5 in)
Gas supply connection 1/4 NPT female
These pneumatic comparison test pumps are designed for testing pressure measuring instruments
against master test instruments. These cost effective instruments provide precise control for impor-
tant calibration requirements and include many features found in our popular line of pneumatic
deadweight testers.
Pneumatic comparison test pumps
Model P5510
Dual pressure/
vacuum capability
Bench mount design
Built-in handpump
as pressure/vacuum
source
High quality needle
valve for fine control
Test port adapt-
ers which require
no PTFE tape or
wrenches
Model P5513
High pressure
pneumatic
operation
Screw press for
fine pressure
adjustments
High quality
needle valves
for fine control
Test port
adapters which
require no PTFE tape or wrenches
Sturdy carrying case with lid
Note: P5513 requires a pressure source, such as a nitrogen bottle, for operation.
2700G Reference
Pressure Gauge
Provides best-in-
class 0.02% FS
accuracy
Available in ranges
up to 10,000 psi
(70 MPa)
Can be used with
P5500 Series Com-
parison Test Pumps
for a complete pressure calibration solution
Note: Detailed product information about the 2700G, including a product data