Instructions

Table Of Contents
This command may be password protected (see Section 8.2.10, Password
Commands).
8.2.1.26 ALARm:RHUMidity<chn>:RATE:LIMit? [MIN|MAX|DEF]
Returns the humidity rate alarm limit for the specified channel in %RH per
hour.
Example: ALAR:RHUM1:RATE:LIM?
Response: 5
The RHUMidity suffix, <chn>, specifies the channel (1 or 2). Specifying the
MIN, MAX, or DEF parameter returns the minimum, maximum, or default
value respectively.
8.2.1.27 ALARm:RHUMidity<chn>:RATE:LIMit <float>|MIN|MAX|DEF
Sets the humidity rate alarm limit for the specified channel in %RH per hour.
Example: ALAR:RHUM1:RATE:LIM 1
The RHUMidity suffix, <chn>, specifies the channel (1 or 2). The <float> pa-
rameter specifies the value for the humidity rate alarm limit. Specifying the
MIN, MAX, or DEF parameter sets the humidity rate alarm limit to the mini-
mum, maximum, or default value respectively. This command may be pass-
word protected (see Section 8.2.10, Password Commands).
8.2.1.28 ALARm:RHUMidity<chn>:SENSor?
Indicates whether a humidity sensor alarm event has occurred for the specified
channel.
Example: ALAR:RHUM1:SENS?
Response: 1
The RHUMidity suffix, <chn>, specifies the channel (1 or 2). A value of 1 is
returned if a humidity sensor alarm event has occurred on the specified chan
-
nel. A value of 0 is returned if a humidity sensor alarm event has not occurred
on the specified channel.
8.2.1.29 ALARm:RHUMidity<chn>:SENSor:ENABle?
Returns the state of the humidity sensor alarm enable for the specified channel.
Example: ALAR:RHUM1:SENS:ENAB?
Response: 0
The RHUMidity suffix, <chn>, specifies the channel (1 or 2). A value of 1 is
returned if the humidity sensor alarm is enabled. A value of 0 is returned if the
humidity sensor alarm is disabled.
85
8 Digital Communications Interface
Commands