8.2.1.11 ALARm:PORT?
Returns the alarm port state.
Example: ALAR:PORT:ENAB?
Response: 0
A value of 1 is returned if the alarm port is active (12V), 0 otherwise (0V).
8.2.1.12 ALARm:PORT <bool>
Sets the alarm port state.
Example: ALAR:PORT:ENAB 1
The <bool> parameter activates (1 or ON) or deactivates (0 or OFF) the alarm
port, overriding the state determined by alarms. This command may be pass
-
word protected (see Section 8.2.10, Password Commands).
8.2.1.13 ALARm:PORT:ENABle?
Returns the alarm port enable setting.
Example: ALAR:PORT:ENAB?
Response: 0
A value of 1 is returned if the alarm port is enabled. A value of 0 is returned if
the alarm port is disabled.
8.2.1.14 ALARm:PORT:ENABle <bool>
Sets the alarm port enable setting.
Example: ALAR:PORT:ENAB 1
The <bool> parameter turns the alarm port enable on (1 or ON) or off (0 or
OFF). This command may be password protected (see Section 8.2.10, Pass
-
word Commands).
8.2.1.15 ALARm:POWer?
Indicates whether a power loss alarm event has occurred.
Example: ALAR:POW?
Response: 0
A value of 1 is returned if a power loss alarm event has occurred. A value of 0
is returned if a power loss alarm event has not occurred.
8.2.1.16 ALARm:POWer:ENABle?
Returns the state of the power loss alarm enable.
Example: ALAR:POW:ENAB?
1620A “DewK” Thermo-hygrometer
User’s Guide
82