Instructions

Table Of Contents
A value of 1 is returned if the sensor settings password-protection is enabled.
Otherwise, a value of 0 is returned.
8.2.10.12 SYSTem:PASSword:SENSor <bool>
Sets the state of the sensor settings password-protection.
Example: SYST:PASS:SENS 1
The <bool> parameter turns the sensor settings password-protection on (1 or
ON) or off (0 or OFF). This command is password protected. Commands pro
-
tected with this setting are: CALC:PAR:DATE:TIME, CALC:PAR:RES,
ROUT:CLOS, ROUT:OPEN, SENS:AVER, SENS:IDEN, SENS:LOCK,and
TRIG:TIME.
8.2.10.13 SYSTem:PASSword:STATistics?
Returns the state of the statistics settings password-protection.
Example: SYST:PASS:STAT?
Response: 0
A value of 1 is returned if the statistics settings password-protection is enabled,
otherwise, a value of 0 is returned.
8.2.10.14 SYSTem:PASSword:STATistics <bool>
Sets the state of the statistics settings password-protection.
Example: SYST:PASS:STAT 1
The <bool> parameter turns the statistics settings password-protection on (1 or
ON) or off (0 or OFF). This command is password protected. Commands pro
-
tected with this setting are: CALC:AVER:CLE and DATA:DST... ,
8.2.10.15 SYSTem:PASSword:TIME?
Returns the state of the date and time password-protection.
Example: SYST:PASS:TIME?
Response: 0
A value of 1 is returned if the date and time password-protection is enabled.
otherwise, a value of 0 is returned.
8.2.10.16 SYSTem:PASSword:TIME <bool>
Sets the state of the date and time password-protection.
Example: SYST:PASS:TIME 1
The <bool> parameter turns the date and time password-protection on (1 or
ON) or off (0 or OFF). This command is password protected. Commands pro
-
1620A “DewK” Thermo-hygrometer
User’s Guide
120