8.2.6.1 CALibrate<chn>:ALERt?
Returns the state of the sensor calibration expired alert enable for the specified
channel.
Example: CAL1:ALER?
Response: 0
The CALibrate suffix, <chn>, specifies the channel (1 or 2). A value of 1 is re
-
turned if the calibration alert is enabled for the specified channel. A value of 0
is returned if the calibration alert is disabled for the specified channel.
8.2.6.2 CALibrate<chn>:ALERt <bool>
Sets the state of the sensor alert enable for the specified channel.
Example: CAL1:ALER 1
The CALibrate suffix, <chn>, specifies the channel (1 or 2). The <bool> pa
-
rameter turns the sensor alert on (1 or ON) or off (0 or OFF). This command is
password protected (see Section 8.2.10, Password Commands).
8.2.6.3 CALibrate<chn>:DATE:CALibrate?
Returns the sensor calibration date for the specified channel.
Example: CAL1:DATE:CAL?
Response: 2003,9,17
The CALibrate suffix, <chn>, specifies the channel (1 or 2). The response is re-
turned in the format, <year>,<month>,<day>.
8.2.6.4 CALibrate<chn>:DATE:CALibrate (<year>,<month>,<day>)
Sets the sensor calibration date for the specified channel.
Example: CAL1:DATE:CAL 2003,9,17
The CALibrate suffix, <chn>, specifies the channel (1 or 2). The <year>,
<month>, and <day> parameters must all be specified. The <year> parameter is
a four-digit number. The <month> parameter is a one or two-digit number (1 to
12). The <day> parameter is a one or two-digit number (1 to 31). This com
-
mand is password protected (see Section 8.2.10, Password Commands).
8.2.6.5 CALibrate<chn>:DATE:DUE?
Returns the sensor calibration due date for the specified channel.
Example: CAL1:DATE:DUE?
Response: 2004,9,17
The CALibrate suffix, <chn>, specifies the channel (1 or 2). The response is re
-
turned in the format, <year>,<month>,<day>.
1620A “DewK” Thermo-hygrometer
User’s Guide
106