Instructions
Table Of Contents
- 1620A “DewK” Thermo-Hygrometer
- 1 Before You Start
- 2 Introduction
- 3 Specifications and EnvironmentalConditions
- 4 Quick Start
- 5 Parts and Controls
- 6 General Operation
- 7 Menu Functions
- 8 Digital Communications Interface
- 9 Calibration Procedure
- 10 Maintenance
- 11 Troubleshooting
- Index

Response: 25.576,29.30
or
Response: 1,1,25.629,C,29.32,%,2003,9,16,11,2,38
Example 2: MEAS?
Response: 25.582,29.32,26.341,37.96
or
Response: 1,1,25.629,C,29.29,%,2,0,C,0,%,2003,9,16,11,1,42
This command is equivalent to the FETC? command.
8.2.2.17 READ? [<chn>]
Returns a comma delimited list of the most recent measurements for the speci-
fied channel or for both channels.
Example 1: READ? 1
Response: 25.576,29.30
or
Response: 1,1,25.629,C,29.32,%,2003,9,16,11,2,38
Example 2: READ?
Response: 25.582,29.32,26.341,37.96
or
Response: 1,1,25.629,C,29.29,%,2,0,C,0,%,2003,9,16,11,1,42
This command is equivalent to the FETC? command.
8.2.3 Data Commands
The data commands are used to access daily statistics or recorded
measurements.
8.2.3.1 DATa:DSTatistics:ENABle?
Returns the state of the daily statistics recording enable.
Example: DAT:DST:ENAB?
Response: 0
A value of 1 is returned if the daily statistics recording is enabled. A value of 0
is returned if the daily statistics recording is disabled.
8.2.3.2 DATa:DSTatistics:ENABle <bool>
Sets the state of the daily statistics recording.
1620A “DewK” Thermo-hygrometer
User’s Guide
96