Instructions
  Front-Panel Operation 
  Configuring the Remote Interface  3 
  3-25 
USB. If STOP is selected during storage to the USB device, readings in internal memory 
will be written directly to the device. Note that this may take several seconds. A USB 
memory device should never be removed while a write is occurring. 
The data is initially stored in internal memory and then transferred to the USB memory 
device. For large data samples (i.e. > 10,000 readings), when 10,000 readings have been 
stored in the internal memory, the data will be transferred to the USB memory device. 
The front panel display shows “BUSY WRITING USB" during the write. For 
10,000 samples, the write time will typically take ~14 seconds. 
In ACV, you can press dB, dBm, or zero while data is being stored. Units (and dgC, dgF, 
K etc) can be changed in the middle of a USB readings file as well. 
Each *.CSV file stored or copied to the USB memory device shows the date + time of the 
first sample line/row and the date + time of the last sample in each file. 
Note 
Readings are not stored during the time each file is written to USB memory 
stick, which can take 7 - 15 seconds. As a result, when storing more than 
10,000 readings, there will be data gaps during the time the file is written 
to the USB memory device. 
Note 
When using Mx+B and other math functions, the internal sampling rate is 
slowed to permit data collection without loss. As an example, with Mx+B 
active, the maximum dcV storage rate is ~340 reading/s. 
USB Memory Device Compatibility and Special Instruction 
After inserting a USB memory device (Mass Storage Device) into the 8846A, wait 
5+ seconds for the USB memory to be mounted before initiating any memory or 
measurement activity. 
Press [MEMORY], MANAGE MEMORY, and USB STATUS to verify a USB 
memory device is readable by the 8846A Not all USB memory devices are usable. 
Devices that try to upload their own driver are generally not compatible. 
Wait 3 or more seconds after the USB memory device light stops indicating any activity 
before removing the USB memory device. Powering the 8846A down or removing the 
USB memory device prematurely can leave the USB memory device unreadable. 
Configuring the Remote Interface 
Selecting an interface port, setting up the ports, and selecting the command set the Meter 
will recognize is accomplished through the Instrument Setup key. For information on the 
commands that will remotely control the Meter, refer to the Programmers Manual. 
Checking the Meter’s Calibration Date 
To read the Meter’s calibration date: 
1.  Press I 
2.  Press the CAL soft key as shown below. 