Controls and Indicators 
Front-Panel Feature Descriptions 
Table 3-1 shows the Meter’s front-panel controls and connectors. 
Table 3-1. Front-Panel Controls and Connectors 
8846A 
6-1/2 DIGIT PRECISION MULTIMETER
DC V
 ACV
DC I ACI
TRIG
FREQ
PERIOD
TEMP
ZERO
INSTR
SETUP
MEAS
SETUP
RANGE
F1 F2 F3 F4 F5
BACK
MEMORY
ANALYZE
INPUT SENSE
HI
FRONT
 REAR
10 A
1V
400
mA
4
W
HI
2W/4W
V
300V
FUSED
1000V CAT I
 600V CAT II
12
1
16
15
2 3 4
68101113
14
7 59
PERIOD
FREQ
8845A Only
caw04.eps
Item Description 
A 
Input HI and LO connectors. Input connectors for Volts, 2-wire Ohms, Hz, Period, 
Temperature, and Capacitance measurements. Input connectors source current 
for 4-wire Ohms measurements. All measurements use the Input LO connector as 
a common input. The LO input is isolated, and may be safely floated up to 1000 V 
above earth ground regardless of the measurement type. 1000 V is the maximum 
voltage rating between the Input HI and LO connectors and between each HI and 
LO input and earth ground. 
B 
Sense HI and LO connectors. Sense connectors sense the voltage across the 
unknown resistance for 4-wire Ohms measurements or provide the DCV 
reference Input for DCV Ratio measurements. 
C 
Soft keys F1 through F5. Soft keys are used to select various menu options while 
navigating the Meter’s menus. Each soft key’s function is identified with a label in 
the bottom row of the display. Keys without a label above them are inactive 