Instructions
  Making Measurements 
  Testing Continuity  4 
  4-15 
- 246.645
º
F
F1 F2 F3 F4 F5
caw11f.eps 
To change the temperature scale, refer to the “Setting the Default Temperature Scale” 
section in Chapter 3 of this manual. Available scales are Celsius, Fahrenheit, and Kelvin. 
Refer to the “Adjusting the Meter’s Range” section in Chapter 3 of this manual for 
information on how to adjust the measurement range.  
Function modifiers: 
4Wire  Switches the measurement input connectors to 4-wire measurement 
for 4-wire RTDs. 4-wire RTDs make more accurate measurements. 
RTD 385  Default RTD type. All coefficients are predefined. 
RO  Used to select a different value of RTD resistance at 0 °C. 
ALPHA  Used to set the first coefficient of the Calendar-Van Dusen equation. 
2ND MEA  Cycles the secondary display through the measurement functions 
listed below and then off. 
OHMS - Displays the resistance of the RTD. 2-wire resistance is used 
when in 2-wire temperature mode and 4-wire resistance when the 
Meter is in 4-wire RTD mode. 
Testing Continuity 
Continuity testing determines whether a circuit is intact (i.e., has a resistance less than the 
threshold). The threshold is selectable between 1 and 1000 Ω. 
To perform a continuity test: 
1.  Press S on the 8846A or Q on the 8845A. An example of the continuity display 
is shown below. 
OPEN
F1 F2 F3 F4 F5
caw12f.eps 
Connect the test leads as shown in Figure 4-1. 
Note 
The Beeper cannot be disabled in continuity. Pressing BEEPER OFF turns 
off the beeper for errors. 
To set the threshold value, refer to the “Setting the Continuity Threshold Resistance” 
section in Chapter 3 of this manual. 