Instructions
  Making Measurements 
  Measuring Resistance  4 
  4-9 
0.217384
F1 F2 F3 F4 F5
caw031.eps 
3.  If not already highlighted as shown above, press the 4WIRE soft key to switch to a 
four-wire measurement. 
2
INPUT SENSE
HI
LO
FRONT
 REAR
10 A
1V
400
mA
4
W
HI
LO
2W/4W
V
300V
FUSED
1000 V CAT I
 600V CAT II
caw023.eps 
Figure 4-2. Input Connections for 4-wire Resistance Measurements 
To make a four-wire resistance measurement using Fluke’s 2X4 test leads: 
1.  Connect the test leads to the Meter’s input connectors as show in Figure 4-3. 
2.  Press N. 
3.  If not already highlighted, press the 2X4WIRE soft key. 