Instructions
8845A/8846A 
Users Manual 
3-20 
Note 
Each memory file will store up to 10000 readings. If the number of samples 
is set to a value greater than 10000, consecutive memory files will be used 
to store all the samples. If the readings fill the last file (999), reading 
storage will stop. 
Recalling Readings from Memory 
To recall readings from internal memory: 
1.  Press M. 
2.  Press the RECALL READING soft key. 
The Meter displays the first stored reading from the internal file. Four soft keys 
provide the means to scroll through the readings stored in the file. The FIRST soft 
key displays the first reading in the file while the LAST soft key displays the last 
reading. Use the <-- and --> soft keys to move forward and backward through the 
file one reading at a time. 
To recall readings from external memory (8846A Only), the memory device must be 
removed from the 8846A and plugged into a PC where the comma delimited files can be 
read. Each file is labeled MEAS0XXX.CSV. The XXX is the file number starting at 001 
and going up to 999. Each file is date and time stamped. 
Storing Meter Configuration Information 
Up to five meter configurations can be stored in the Meter’s internal memory. The 
8846A, with optional USB memory installed, can store an additional 99 configurations in 
the external memory. 
To store a meter configuration in the Meter’s internal memory: 
1.  Press M. 
2.  Press the STORE CONFIG soft key as shown below. 
F1 F2 F3 F4 F5
 0.5993
+
caw033.eps 
3.  Press the STORE INT MEM soft key. 
4.  Press one of the five soft keys labeled with the desired memory location to store the 
present Meter configuration. 
To store a meter configuration in to optional external memory (8846A Only): 
1.  Press M. 
2.  Press the STORE CONFIG soft key. 
3.  Press the STORE USB soft key. 
The Meter labels the first four soft keys with the first four memory locations. 
CONFIG01 through CONFIG04. The fifth soft key is labeled MORE to allow 
access to all 100 memory locations. 