Instructions
  Introduction and Specifications 
  Electrical Specifications  1 
  1-17 
Accuracy (8846A) 
Accuracy is given as ± (% measurement + % of range) 
Range 
24 Hour 
(23 ±1 °C) 
90 Days 
(23 ±5 °C) 
1 Year 
(23 ±5 °C) 
Temperature 
Coefficient/ °C 
Outside 18 to 28 °C 
100 μA  0.01 + 0.02  0.04 + 0.025  0.05 + 0.025  0.002 + 0.003 
1 mA  0.007 + 0.005  0.030 + 0.005  0.05 + 0.005  0.002 + 0.0005 
10 mA  0.007 + 0.02  0.03 + 0.02  0.05 + 0.02  0.002 + 0.002 
100 mA  0.01 + 0.004  0.03 + 0.005  0.05 + 0.005  0.002 + 0.0005 
400 mA 
[3]
  0.03 + 0.004  0.04 + 0.005  0.05 + 0.005  0.005 + 0.0005 
1 A 
[2]
  0.03 + 0.02  0.04 + 0.02  0.05 + 0.02  0.005 + 0.001 
3 A 
[1][2]
  0.05 + 0.02  0.08 + 0.02  0.1 + 0.02  0.005 + 0.002 
10 A 
[2]
  0.1 + 0.008  0.12 + 0.008  0.15 + 0.008  0.005 + 0.0008 
[1] Part of 10 A range 
[2] Available at front panel connectors only 
[3] 400 mA available in software version 2.0 or greater only. 400 mA continuously; 550 mA for 2 minutes on, 1 
minute off. 
Accuracy (8845A) 
Accuracy is given as ± (% measurement + % of range) 
Range 
24 Hour 
(23 ±1 °C) 
90 Days 
(23 ±5 °C) 
1 Year 
(23 ±5 °C) 
Temperature 
Coefficient/ °C 
Outside 18 to 28 °C 
100 μA  0.01 + 0.02  0.04 + 0.025  0.05 + 0.025  0.002 + 0.003 
1 mA  0.007 + 0.005  0.030 + 0.005  0.05 + 0.005  0.002 + 0.0005 
10 mA  0.007 + 0.02  0.03 + 0.02  0.05 + 0.02  0.002 + 0.002 
100 mA  0.01 + 0.004  0.03 + 0.005  0.05 + 0.005  0.002 + 0.0005 
400 mA 
[3]
  0.03 + 0.004  0.04 + 0.005  0.05 + 0.005  0.005 + 0.0005 
1 A 
[2]
  0.03 + 0.02  0.04 + 0.02  0.05 + 0.02  0.005 + 0.001 
3 A 
[1][2]
  0.05 + 0.02  0.08 + 0.02  0.10 + 0.02  0.005 + 0.002 
10 A 
[2]
  0.10 + 0.008  0.12 + 0.008  0.15 + 0.008  0.005 + 0.0008 
[1] Part of 10 A range 
[2] Available at front panel connectors only 
[3] 400 mA available in software version 2.0 or greater only. 400 mA continuously; 550 mA for 2 minutes on, 1 
minute off. 
Additional Current Errors 
Digits NPLC 
Additional NPLC Noise Error for 
1 mA, 100 mA, 400 mA, 3 A and 
10 A 
Additional NPLC Noise Error 
for 100 μA, 10 mA, 1 A 
6½  100  0 % of range  0 % of range 
6½  10  0 % of range  0 % of range 
5½  1  0.001 % of range  0.01 % of range 
5½ 0.2 
0.011 % of range ±4 μA  0.11 % of range ±4 μA 
4½ 0.02 
0.04 % of range ±4 μA  0.28 % of range ±4 μA 
AC Current 
The following ac current specifications are for sinusoidal signals with amplitudes greater than 5 % of range. For inputs 
from 1 % to 5 % of range, add an additional error of 0.1 % of range. 
Input Protection....................................................... Tool accessible 11 A/1000 V and 440 mA/1000 V fuses, limits of 
400 mA continuous 550 mA for 2 minutes on, 1 minute off. 
Measurement Method ............................................. ac-coupled true-rms, dc-coupled to the fuse and shunt (no blocking 
capacitor) 
AC Filter Bandwidth 
Slow .................................................................... 3 Hz to 10 kHz 
Medium ............................................................... 20 Hz to 10 kHz 