User manual

33
Digital Communication Interface
Interface Commands
Command Description
Command
Format
Command
Example
Returned
Returned
Example
Acceptable
Values
Serial Interface Menu
Read serial sample setting sa[mple] sa sa: 99999 sa: 1
Set serial sampling setting to n seconds sa[mple]=n sa=0 0 to 10,000
Set serial duplex mode: du[plex]=f[ull]/
h[alf]
FULL or HALF
Set serial duplex mode to full du[plex]=f[ull] du=f
Set serial duplex mode to half du[plex]=h[alf] du=h
Set serial linefeed mode: lf[eed]=on/of[f] ON or OFF
Set serial linefeed mode to on lf[eed]=on lf=on
Set serial linefeed mode to off lf[eed]=of[f] lf=of
Calibration Menu
Read R0 calibration parameter r[0] r r0: 999.999 r0: 100.7
Set R0 calibration parameter to n r[0]=n r=100.7 95.0 to 105.0
Read ALPHA calibration parameter al[pha] al al: 9.99999999 al: 0.003865
Set ALPHA calibration parameter to n al[pha]=n al=0.003865 0.002 to 0.006
Read DELTA calibration parameter de[lta] de de: 1.50
Set DELTA calibration parameter de[lta]=n de=1.37 de: 9.99999 0.0–3.0
Miscellaneous Other Commands
Read rmware version number *ver[sion] *ver ver.9999x,9.99 ver.9102S,1.10
Read structure of all commands h[elp] h list of
commands
Read ALL operating parameters all all list of
parameters
Read set-point *sr *sr 999.999 ohms 100.123 ohms
Legend: [] Optional command data
{} Returns either information
n Numeric data supplied by user
9 Numeric data returned to user
x Character data returned to user
Note: When DUPLEX is set to FULL and a command is sent to READ, the command is returned
followed by a carriage return and linefeed. Then the value is returned as indicated in the
RETURNED column.