Controller Operation
Calibration Parameters
To change the ALPHA setting, press the up or down arrows.
38600 NewALPHAsetting
To accept the new setting, press “SET”. Press “EXIT” to continue without storing the
new value.
S
AcceptthenewALPHAsetting
6.12.4 DELTA
This probe parameter characterizes the curvature of the resistance-temperature
relationship of the sensor. The value of this parameter is set at the factory for best
instrument accuracy.
DeLta FlashesforonesecondandthentheDELTAsettingis
displayed
0.0000 CurrentDELTAsetting
To change the DELTA setting, press the up or down arrows.
0.1000 NewDELTAsetting
To accept the new setting, press “SET”. Press “EXIT” to continue without storing the
new value.
S
AcceptthenewDELTAsetting