9100S Dry-Well Calibrator Users Guide 9100S___ugeng0000
Limited Warranty & Limitation of Liability Each product from Fluke Corporation, (“Fluke”) is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is one year for Dry-Well Calibrators. The warranty period begins on the date of the shipment. Parts, product repairs, and services are warranted for 90 days.
Table of Contents 1 Before You Start........................................................................1 1.1 1.2 1.3 Introduction................................................................................................ 1 Symbols Used............................................................................................ 1 Safety Information...................................................................................... 3 1.3.1 1.3.2 1.4 2 4 Unpacking........................
9100S Dry-Well Calibrator 6.3.1 6.3.2 Scan Control...................................................................................................22 Scan Rate.......................................................................................................22 6.4 6.5 6.6 Set-point Resistance................................................................................ 22 Secondary Menu..................................................................................... 23 Heater Power........
11 Troubleshooting......................................................................43 11.1 Troubleshooting Problems, Possible Causes, and Solutions................... 43 11.2 CE Comments.......................................................................................... 44 11.2.1 EMC Directive................................................................................................44 11.2.2 Low Voltage Directive (Safety)..................................................................
9100S Dry-Well Calibrator Tables Table 1 International Electrical Symbols............................................................. 1 Table 2 Controller communications commands...............................................
Figures Figure 1 Back Panel.......................................................................................... 13 Figure 2 Side view of the 9100S showing the prop stand................................ 14 Figure 3 Front panel.......................................................................................... 15 Figure 4 Fixed block options............................................................................ 16 Figure 5 Controller operation flow chart..................................
Before You Start Symbols Used 1 Before You Start 1.1 Introduction The Fluke 9100S Mid-Range Field Calibrator may be used as a portable instrument or bench top temperature calibrator for calibrating thermocouple and RTD temperature probes. This instrument is small enough to use in the field, and accurate enough to use in the lab. Calibrations may be done over a range of 35°C to 375°C (95°F to 707°F). Temperature display and setability resolution is 0.1 degrees.
9100S Dry-Well Calibrator Symbols Used Symbol Description AC-DC Battery Complies with European Union directives DC Double Insulated Electric Shock Fuse PE Ground Hot Surface (Burn Hazard) Read the User’s Guide (Important Information) Off On Canadian Standards Association CAT II OVERVOLTAGE (Installation) CATEGORY II, Ploution Degree 2 per IEC1010-1 refers to the level of Impulsive Withstand Voltage protection provided.
Before You Start Safety Information 1.3 Safety Information Use this instrument only as specified in this manual. Otherwise, the protection provided by the instrument may be impaired. Refer to the safety information in the Warnings and Cautions sections below. The following definitions apply to the terms “Warning” and “Caution”. ●● ●● 1.3.1 “Warning” identifies conditions and actions that may pose hazards to the user. “Caution” identifies conditions and actions that may damage the instrument being used.
9100S Dry-Well Calibrator Safety Information BURN HAZARDS ●● DO NOT turn the instrument upside down with the inserts in place; the inserts will fall out. ●● DO NOT operate near flammable materials. ●● Use of this instrument at HIGH TEMPERATURES for extended periods of time requires caution. ●● DO NOT touch the well access surface of the instrument. ●● The block vent may be very hot due to the fan blowing across the heater block of the dry-well.
Before You Start Authorized Service Centers ●● ●● ●● ●● ●● ●● ●● ●● ●● ●● ●● ●● ●● 1.4 cm) of clearance around the instrument. Component lifetime can be shortened by continuous high temperature operation. DO NOT use fluids to clean out the well. Never introduce any foreign material into the probe hole of the insert. Fluids, etc. can leak into the instrument causing damage. DO NOT change the values of the calibration constants from the factory set values.
9100S Dry-Well Calibrator Authorized Service Centers Phone: +1.801.763.1600 Telefax: +1.801.763.1010 E-mail: support@flukecal.com Fluke Nederland B.V. Customer Support Services Science Park Eindhoven 5108 5692 EC Son NETHERLANDS Phone: +31-402-675300 Telefax: +31-402-675321 E-mail: ServiceDesk@fluke.
Before You Start Authorized Service Centers E-mail: antng@singa.fluke.
Specifications and Environmental Conditions Environmental Conditions 2 Specifications and Environmental Conditions 2.1 2.2 Specifications Range 35°C to 375°C (95°F to 707°F) Accuracy ±0.25°C at 50°C ±0.25°C at 100°C ±0.5°C at 375°C Stability ±0.07°C at 50°C ±0.1°C at 100°C ±0.3°C at 375°C Resolution 0.1°C or °F Well-to-Well Uniformity ±0.2°C with sensors of similar size at equal depths within wells Heating Times 35 to 375°C: 9.
Quick Start Setting the Temperature 3 Quick Start 3.1 Unpacking Unpack the dry-well carefully and inspect it for any damage that may have occurred during shipment. If there is shipping damage, notify the carrier immediately. Verify that the following components are present: ●● ●● ●● ●● ●● 3.
3. Press “SET” to program in the new set-point. 4. Press and hold “EXIT” to return to the temperature display. When the set-point temperature is changed the controller switches the well heater on or off to raise or lower the temperature. The displayed well temperature gradually changes until it reaches the set-point temperature. The well may require 5 to 10 minutes to reach the set-point depending on the span. Another 5 to 10 minutes is required to stabilize within ±0.1°C of the set-point.
Parts and Controls Rear Panel 4 Parts and Controls The user should become familiar with the dry-well calibrator and its parts: (See Figures 1, 2, and 3). 4.1 Rear Panel Power Cord - The removable power cord, (Figure 1 on this page) attaches to the back side of the instrument. It plugs into a standard 115 VAC (optional 230 VAC) grounded socket. Figure 1 Back Panel Power Switch - The power switch is located on the back panel of the instrument. The switch is either on or off.
9100S Dry-Well Calibrator Side View RS-232 - The RS-232 serial port provides a means for connecting the instrument to a computer or a printer using the included serial cable. WARNING: Always leave enough clearance in front of the calibrator to allow for safe and easy installation and removal of probes. 4.2 Side View Prop Stand - The prop stand (Figure 2 on this page) is located on the bottom side of the instrument and lays flat against the bottom of the instrument when not in use.
Parts and Controls Constant Temperature Block Assembly 4.3 Front Panel Well Block - Located on the right side of the front panel is the well opening where a block (Figure ) designed to accept different probe sizes is installed. Each block is designed to accept different diameter temperature sensors. Four different blocks are available. See Section , Constant Temperature Block Assembly for details.
9100S Dry-Well Calibrator Constant Temperature Block Assembly the temperature of the block. The entire assembly is suspended in an air cooled chamber thermally isolated from the chassis and electronics. WARNING: The opening in front of the block may be very hot due to air blowing forward. The constant temperature block is available in the following configurations.
General Operation Changing Display Units 5 General Operation 5.1 Setting the Temperature Section explains in detail how to set the temperature set-point on the calibrator using the front panel keys. The procedure is summarized here. 1. 2. 3. 4. Press “SET” twice to access the set-point value. Press u or d to change the set-point value. Press “SET” to program in the new set-point. Press and hold “EXIT” to return to the temperature display.
Controller Operation Temperature Set-point 6 Controller Operation This chapter discusses in detail how to operate the dry-well temperature controller using the front control panel. Using the front panel key-switches and LED display the user may monitor the well temperature, set the temperature set-point in degrees C or F, monitor the heater output power, adjust the controller proportional band, and program the calibration parameters, operating parameters, and serial interface configuration.
1 100. Set-point memory 1 location, 100°C currently used To change to another set-point memory press the up or down arrow. 4 300. New set-point memory 4 location, 300°C Press “SET” to accept the new selection and access the set-point value. Press “EXIT” to continue and to ignore any changes. S 6.2.2 Accept selected set-point memory Set-point Value The set-point value may be adjusted after selecting the set-point memory and pressing “SET”. 4 200.
Controller Operation Scan SET Display Temperature + Units °C/°F u SET Select Setpoint + DOWN SET SET/EXIT Set-Point Resistance Adjust Setpoint Secondary Functions SET/EXIT SET Units °C/°F + EXIT SET/EXIT Display Power Scan On/Off SET/EXIT SET/EXIT Set Proportional Band Scan Rate SET SET/EXIT EXIT Configuration Menu SET EXIT UP DOWN Serial Interface Menu SET SET HL BAUD Rate Adj.
9100S Dry-Well Calibrator Set-point Resistance 6.3.1 Scan Control The scan is controlled with the scan on/off function that appears in the main menu after the temperature scale units. Sc Sc flashes for one second and then the current scan setting Off Scan function off is displayed Press the up or down arrow to toggle the scan on or off. On Scan function on Press “SET” to accept the present setting and continue. S 6.3.2 Accept scan setting Scan Rate The scan rate can be set from 0.1 to 99.
Controller Operation Proportional Band 99. Whole number portion of the set-point resistance flashes for two seconds and then the fraction portion of the current set-point resistance setting is displayed .222 Fraction portion of the current set-point resistance setting The set-point resistance is 99.222. 6.5 Secondary Menu Functions which are used less often are accessed within the secondary menu. The secondary menu is accessed by pressing “SET” and “EXIT” simultaneously and then releasing.
9100S Dry-Well Calibrator Controller Configuration to lower the temperature back down. In this way the temperature is maintained at a fairly constant level. The temperature stability of the well and response time depend on the width of the proportional band. If the band is too wide, the well temperature deviates excessively from the set-point due to varying external conditions.
Controller Operation Serial Interface Parameters parameters, serial interface parameters, and calibration parameters. The menus are selected using the up and down arrows and then pressing “SET”. 6.9 Operating Parameters The operating parameters menu is indicated by, PAr Operating parameters menu The operating parameters menu contains the High Limit parameter. 6.10 High Limit The High Limit parameter adjusts the upper set-point temperature. The factory default and maximum are set to 375°C (707°F).
9100S Dry-Well Calibrator Serial Interface Parameters bAUd Flashes bAUd for one second and then displays the setting 2400b Current baud rate The baud rate of the serial communications may be programmed to 300, 600, 1200, 2400 (default), 4800, or 9600 baud. Use the up or down arrows to change the baud rate value. 4800 b New baud rate Press “SET” to accept the new setting or “EXIT” to abort the operation and skip to the next parameter in the menu. 6.11.
Controller Operation Calibration Parameters HALF New duplex mode setting Press “SET” to accept the new setting or “EXIT” to abort the operation and skip to the next parameter in the menu. 6.11.4 Linefeed The final parameter in the serial interface menu is the linefeed mode. This parameter enables (on) or disables (off) transmission of a linefeed character (LF, ASCII 10) after transmission of any carriage-return.
9100S Dry-Well Calibrator Calibration Parameters new value. Pressing “EXIT” causes the parameter to be skipped ignoring any changes that may have been made. 7.12.1 6.12.1 Hard Cutout This parameter is the temperature above which the unit shuts down automatically. The value of this parameter is set at the factory to approximately 400°C and cannot be changed by the user. The hard cutout parameter is indicated by, CtOut Flashes for one second and then the hard cutout setting is displayed 400.
Controller Operation Calibration Parameters To change the ALPHA setting, press the up or down arrows. 38600 New ALPHA setting To accept the new setting, press “SET”. Press “EXIT” to continue without storing the new value. S 6.12.4 Accept the new ALPHA setting DELTA This probe parameter characterizes the curvature of the resistance-temperature relationship of the sensor. The value of this parameter is set at the factory for best instrument accuracy.
Digital Communication Interface RS-232 Connection 7 Digital Communication Interface This instrument is capable of communicating with and being controlled by other equipment through the digital serial interface. With a digital interface, the instrument may be connected to a computer or other equipment. This allows the user to set the set-point temperature, monitor the temperature, and access any of the other controller functions, all using remote communications equipment.
9100S Dry-Well Calibrator Interface Commands The serial port can be used to transmit measurements to a computer or printer or to change settings of the instrument from a computer. A full list of commands follows in Section . Commands sent to the instrument must end with an EOS character which is a carriage return (CR, ASCII 13) or linefeed character (LF, ASCII 10). Commands can be sent with upper or lower case letters. Data returned from the instrument ends with a carriage return.
Digital Communication Interface Interface Commands Command Description Secondary Menu Read proportional band setting Command Format Command Example Returned Returned Example pb: 999.9 pb: 15.9 pr[op-band] pr Set proportional band to n pr[op-band]=n pr=8.83 Read heater power (duty cycle) Configuration Menu Operating Parameters Menu po[wer] po po: 999.9 po: 6.5 Read High Limit hl[imit] hl hl: 9999.9 hl: 375.
Test Probe Calibration Dry-well Characteristics 8 Test Probe Calibration For optimum accuracy and stability, allow the calibrator to warm up for 10 minutes after power-up and then allow adequate stabilization time after reaching the set-point temperature. After completing operation of the calibrator, allow the well to cool by setting the temperature to 25°C for one-half hour before switching the power off. 8.1 Calibrating a Single Probe Insert the probe to be calibrated into the well of the instrument.
Calibration Procedure Calibration Procedure 9 Calibration Procedure Note: This procedure is to be considered a general guideline. Each laboratory should write their own procedure based on their equipment and their quality program. Each procedure should be accompanied by an uncertainty analysis also based on the laboratory’s equipment and environment. Sometimes the user may want to calibrate the dry-well to improve the temperature set-point accuracy.
9100S Dry-Well Calibrator Calibration Procedure 9.2.1 Compute DELTA A = T3 − T2 B = T2 − T1 T T T T C = 3 1 − 3 − 2 1 − 2 100 100 100 100 T T T T D = 2 1 − 2 − 1 1 − 1 100 100 100 100 E = R3 − T2 F = R2 − T1 delta = AF − BE DE − CF where: T1 and R1 are the measured temperature and set-point resistance at 50.0 °C T2 and R2 are the measured temperature and set-point resistance at 200.
Calibration Procedure Calibration Procedure Program the new values for DELTA (delta), R0 (rzero), and ALPHA (alpha) into the instrument (see Section 6.12, Calibration Parameters on page 27 and Figure 5 on page 21). 9.2.3 Accuracy and Repeatability Check the accuracy of the dry-well at various points over the calibration range. If drywell does not pass specification at all set-points, repeat the Calibration Procedure.
Maintenance 10 Maintenance ●● ●● ●● ●● ●● ●● ●● ●● ●● ●● This instrument has been designed with the utmost care. Ease of operation and simplicity of maintenance have been a central theme in the product development. Therefore, with proper care the instrument should require very little maintenance. Avoid operating the instrument in an oily, wet, dirty, or dusty environment. If the outside of the instrument becomes soiled, it may be wiped clean with a damp cloth and mild detergent.
Troubleshooting Troubleshooting Problems, Possible Causes, and Solutions 11 Troubleshooting This section contains information on troubleshooting, CE Comments, and a wiring diagram. 11.1 Troubleshooting Problems, Possible Causes, and Solutions In the event that the instrument appears to function abnormally, this section may help to find and solve the problem. Several possible problem conditions are described along with likely causes and solutions.
11.2 CE Comments 11.2.1 EMC Directive Fluke’s equipment has been tested to meet the European Electromagnetic Compatibility Directive (EMC Directive, 89/336/EEC). The Declaration of Conformity for your instrument lists the specific standards to which the unit was tested. 11.2.2 Low Voltage Directive (Safety) In order to comply with the European Low Voltage Directive (73/23/EEC), Fluke equipment has been designed to meet the IEC 1010-1 (EN 61010-1) and the IEC 1010-2010 (EN 61010-2-010) standards.