Information
F4F3F2F1
100.00°
C
Unit
Under
Test
Temperature
Source
Control Output
Reference
Probe
Example
T
emperature
Control
Source
7
• Only one temperature
source can be connected at
a time.
• The 1586A allows
you to set the required
stability band to accept a
temperature measurement.
• The 1586A comes with
Fluke DAQ 6.0 software,
which allows you to handle
and visualize your data.
• Easily export data to a
USB stick in .csv format for
analysis on a computer in a
spreadsheet program.
Connect the calibrator to the temperature source with the appropriate
data cable.
Insert the temperature standard and probes to be tested into the
precision temperature source (i.e. a dry-well).
Connect the probes to be tested and the temperature standard to
the calibrator, and enable the channels they are connected to.
Select the test points and the order they are to be executed.
Start the program.
Move on to another job or unrelated task.
Return to collect and analyze your data.
STEP
1
STEP
2
STEP
3
STEP
4
STEP
6
STEP
7
To perform the test:
Additional resources
For more in depth information about
this application check out these videos
and application notes from Fluke.
1586A Automating Temperature Sensor Calibration Video
Automated Temperature Sensor Calibration with the 1586A Super-DAQ
Eliminating Sensor Errors in Loop Calibrations
TECH
TIPS
STEP
5
7