Information

754
DOCUMENTING PROCESS CALIBRATOR
Null
modem
Interface
cable
Dry-well (3.5 mm)
Dry-well
(DB9)
Hart Dry-well cable kit,
P/N 2111088
3.5 mm
interface cable
Fluke Calibration
Fluke Calibration
Fluke Calibration
5
Record the As-Found test
either using Auto-Test, with
the settling time for the dry-
well in the test delay time, or
manually while observing for
stable test temperatures.
After the As-Found test select
Adjust” and select Yes when
prompted to use a dry-well for
adjusting the transmitter in-
put. The applied temperature
and adjustment will adjust the
transmitter input block to out-
put the correct measurement
by adjusting the transmitter
and sensors output together.
After adjusting the input
with sensor, adjust the
transmitter mA output using
“Output Trim.”
After adjustment is completed,
record the post adjustment
of the transmitter As-Left
and errors of the input.
Sensor and output errors
of the transmitter should
be nominalized, resulting
in improved temperature
measurement performance.
Remove the process measure-
ment sensor and install into the
dry-well temperature calibrator.
Connect the mA measure-
ment jacks of the 754 to the
transmitter and connect the
dry-well communication cable
between the 754 and the
dry-well.
Press HART to query the trans-
mitter for its configuration.
Enable Loop Power on the 754
as required.
Press HART again and config-
ure the calibrator for the test;
select “Measure mA, Source
Dry-Well.”
Select As-Found” to configure
for documenting the test. Be
sure to set the delay time long
enough to allow the dry-well
to change temperatures and
stabilize.
Resistance Temperature
Detectors (RTDs) are almost
always more accurate than
thermocouples (TCs). As long
as the temperature measured
is within the range of the
RTDs, they are a better choice
when accuracy matters.
Thermocouples have a wider
temperature range and are
more durable than RTDs.
Thermocouples are a good
choice for applications with
rough service that experience
destructive vibrations and
repeated temperature cycling.
Above ambient, dry-well cali-
brators stabilize at tempera-
ture faster with increasing
temperatures than decreasing
temperatures.
If the stabilization time for
the dry-well is difficult to
estimate, consider selecting
Manual Test” on the 754 and
wait for the temperature to
stabilize before recording the
measurement.
HART smart transmitters with
RTD probe inputs may allow
for entry of the probe’s certi-
cation constants. By inputting
these constants, the sensor
is matched and measurement
system errors are minimized.
To use a dry-well calibrator with a Fluke 754 to automatically test a transmitter
with sensor:
STEP
1
STEP
7
STEP
8
STEP
9
STEP
10
STEP
2
STEP
3
STEP
4
STEP
5
STEP
6
To perform the test:
Additional resources
For more in depth information about
this application check out these videos
and application notes from Fluke.
Testing, troubleshooting, calibrating process temperature
devices webinar
Selecting a Dry-Well Temperature Calibrator
Temperature calibration application note
Eliminating sensor errors in loop calibrations
TECH
TIPS
5