Information

32
6102, 7102, 7103 Micro-Baths
Micro-Baths provide a convenient way to elimi-
nate the problems of fit and immersion when
calibrating short, square, or odd-shaped sensors.
Proprietary controllers make the Fluke 6102, Fluke
7102, and Fluke 7103 Micro-Baths extremely
stable. They are lighter and smaller than most
dry-wells, have a spill-proof lid, and a convenient
carrying handle, so you can easily take them
where you need to go without a cart and without
removing the fluid.
Key features
Perfect size for calibrating on site
Calibrates sensors of any size
or shape
Stability to ±0.015 °C
Ranges from –30 °C to 200 °C
Spill-proof lid
Convenient handle
Summary specifications
48 mm (1.9 in) diameter,
140 mm (5.5 in) deep tank
6102: 35 °C to 200 °C, 10 lb with fluid
7102: –5 °C to 125 °C, 15 lb with fluid
7103:–30 °C to 125 °C, 22 lb with fluid
Application coverage
RTD and thermocouple calibration
Temperature switch calibration
Liquid-in-glass and bimetallic thermometer cali-
bration
4180 and 4181
Precision Infrared Calibrators
The Fluke Calibra-
tion 4180 and 4181
Precision Infrared
Calibrators give
more consistent,
accurate and reliable
calibrations because
emissivity is radio-
metrically calibrated, and the target size minimizes
size of source effect errors. In addition, these
infrared calibrators simplify calibration because
they uniquely compensate for errors caused by
thermometer emissivity settings.
Key features:
Calibrated radiometrically for meaningful, con-
sistent results
Accredited calibration included
Accurate, reliable performance from –15 °C to
500 °C
Large target size of 152 mm (6 in) is required for
calibrating most thermometers
Light instrument weight of 8.6 kg (19 lbs) makes
it easy to lift and carry
Intuitive, easy-to-read display that indicates
when temperature is stable
Summary specifications
4180 15 °C to 120 °C
± 0.40 °C at 0 °C
4181 35 °C to 500 °C
± 0.70 °C at 200 °C
Application coverage
Infrared thermometer calibration
1586A Super-DAQ
Precision Temperature Scanner
The 1586A Super-DAQ
automates calibration when
connected to Fluke Calibra-
tion temperature source. It
is a powerful thermom-
eter readout and the most
accurate and flexible temperature data acquisition
system on the market. It scans and records tem-
perature, dc voltage, dc current, and resistance for
up to 40 input channels and scan speeds as fast
as 10 channels per second.
Key features:
Measure thermocouples, PRTs, thermistors,
dc voltage, dc current, and resistance
Best-in-class temperature measurement
accuracy
Input channels: Up to 40 isolated
universal inputs
Flexible configuration: Internal High-Capacity
Module and/or DAQ-STAQ Multiplexer
Fluke DAQ 6.0 software included to visualize
and handle data.
USB
Summary specifications
PRTs: ±0.005 °C
Thermocouples: ±0.3 °C
(Type K,T,J w/ internal CJC)
Thermistors: ±0.002 °C
Application coverage
Precision reference thermometer for benchtop
sensor calibration
Automate temperature calibration in the shop
or on the factory floor with dry-wells and
Micro-Baths
Multi-channel data logging