User manual
Table Of Contents
- Table of Contents
- 1 Before You Start 1
- 2 Introduction 7
- 3 Specifications and Environment Conditions 9
- 4 Quick Start 13
- 5 Parts and Controls 15
- 6 General Operation 17
- 7 Controller Operation 19
- 7.1 Target Temperature 19
- 7.2 Temperature Set-point 19
- 7.3 Temperature Scale Units 21
- 7.4 Scan 22
- 7.5 Set-point Resistance 23
- 7.6 Temperature Scale Units 23
- 7.7 Secondary Menu 23
- 7.8 Heater Power 23
- 7.9 Proportional Band 24
- 7.10 Controller Configuration 25
- 7.11 Operating Parameters 25
- 7.12 Serial Interface Parameters 26
- 7.13 Calibration Parameters 27
- 8 Digital Communication Interface 29
- 9 Calibration Procedure 35
- 10 Maintenance 39
- 11 Troubleshooting 41
- Figures
- Tables

9133
User’s Guide
20
UP
UP
DOWN DOWN
SET
Operating
Parameters
Menu
SET
SET
Cal
Menu
ALPHA
DE LTA
Adj. R0
DO NOT CHANGE THESE VALUES. SEE MANUAL
DO NOT CHANGE THESE VALUES. SEE MANUAL
Adj. ALPHA
Serial
Interface
Menu
BAUD
Rate
Adjust
BAUD Rate
Sample
Period
Adj. Sample
Period
Duplex
Mode
Adj. Duplex
Mode
Linefeed
Adjust
Linefeed
BETA
Adjust
BETA
EXITEXIT
EXIT
EXIT
EXIT
UP
DOWN
SET
SET
SET
+
+
+
Display Power
Toggles °C / °F
SET
SET
Select Setpoint
Adjust Setpoint
Units °C/°F
Scan On/Off
Scan Rate
Display
Temperature
Configuration Menu
Secondary Functions
X5
HL
Adj.HL
Displays Set-Point Resistance
Set Proportional Band
R0
Adj. DELTA
Press “SET” to step through the menu and
to store the parameter value.
Press “EXIT” briefly to skip a parameter
without storing the parameter value.
Hold “EXIT” ( 2 seconds) to exit the menu
and display the temperature
≈
Menu Legend:
(2 Seconds)
(2 Seconds)
Automatically returns to display
Figure 4 Controller Operation Flow Chart