User manual

Table Of Contents
4 Quick Start
4.1 Unpacking
Unpack the calibrator carefully and inspect it for any damage that may have
occurred during shipment. If there is shipping damage, notify the carrier
immediately.
Verify that the following components are present:
9133 calibrator
report of calibration
power cord
user's guide
serial cable
target cover
9930 software
4.2 Set Up
Place the calibrator on a flat surface with at least 6 inches of free space
around the instrument. The prop may be swung down to raise the front of
the instrument from a horizontal position. Plug the power cord into a
grounded mains outlet. Observe that the nominal voltage corresponds to
that indicated on the back of the calibrator.
Turn on the power to the calibrator by toggling the power switch on. The
fan should begin quietly blowing air through the instrument and the con
-
troller display should illuminate after 3 seconds. After a brief self-test the
controller should begin normal operation. If the unit fails to operate please
check the power connection.
The thermal electric devices will start operating to bring the temperature of
the calibrator to the set-point temperature and the display will begin to
show the actual target temperature.
Caution: Use the target cover at temperatures below 5°C. If ice forms on
the target, the IR probe will not indicate the correct temperature.
4.3 Power
Plug the instrument power cord into a mains outlet of the proper voltage,
frequency, and current capability. Refer to Section 3.1, Specifications for
power details. Turn the instrument on using the rear panel “POWER”
switch. The instrument turns on and begins to heat to the previously pro
-
13
4
Quick Start