User manual
Table Of Contents
- Table of Contents
- 1 Before You Start 1
- 2 Introduction 7
- 3 Specifications and Environment Conditions 9
- 4 Quick Start 13
- 5 Parts and Controls 15
- 6 General Operation 17
- 7 Controller Operation 19
- 7.1 Target Temperature 19
- 7.2 Temperature Set-point 19
- 7.3 Temperature Scale Units 21
- 7.4 Scan 22
- 7.5 Set-point Resistance 23
- 7.6 Temperature Scale Units 23
- 7.7 Secondary Menu 23
- 7.8 Heater Power 23
- 7.9 Proportional Band 24
- 7.10 Controller Configuration 25
- 7.11 Operating Parameters 25
- 7.12 Serial Interface Parameters 26
- 7.13 Calibration Parameters 27
- 8 Digital Communication Interface 29
- 9 Calibration Procedure 35
- 10 Maintenance 39
- 11 Troubleshooting 41
- Figures
- Tables

3.2 Environmental Conditions
Although the instrument has been designed for optimum durability and
trouble-free operation, it must be handled with care. The instrument should
not be operated in an excessively dusty or dirty environment. Maintenance
and cleaning recommendations can be found in the Maintenance section of
this manual.
The instrument operates safely under the following conditions:
•
temperature range: 5–50°C (41–122°F)
•
ambient relative humidity: maximum 80% for temperature <31°C, de
-
creasing linearly to 50% at 40°C
•
pressure: 75kPa–106kPa
•
mains voltage within ±10% of nominal
•
vibrations in the calibration environment should be minimized
•
altitudes less than 2,000 meters
3.3 Warranty
Fluke Hart Scientific (Hart) warrants this product to be free from defects in
material and workmanship under normal use and service for a period as
stated in our current product catalog from the date of shipment. This war
-
9133
User’s Guide
10
-0.2
-0.15
-0.1
-0.05
0
0.05
0.1
0.15
-30 -10 10 30 50 70 90 110 130 150
Reference Probe °C
delta T °C
Figure 1 Temperature Gradient Between Probe and IR Target Surface at Ambient of 30°C










